4 Specifications
All specifications apply to the unit after a temperature stabilization time of 20
minutes over an ambient range of 25°C ± 5°C.
TT-SI 7002 – order number 15175
Bandwidth
25 MHz @ x50 and x200
15 MHz @ x20
Rise Time
14 ns @ x50 and x200
23 ns @ x20
Attenuation
x20, x50, x200
DC-Accuracy
±2%
CMRR
80 dB @ 60 Hz
60 dB @ 100 Hz
50 dB @ 1 MHz
Maximum Input Voltage @ x20
(DC + AC peak)
±70 V
Maximum Input Voltage @ x50
(DC + AC peak)
±175 V
Maximum Input Voltage @ x200
(DC + AC peak)
±700 V
Absolute Maximum Input Voltage
(each side to ground)
600 Vrms
Input Impedance
(differential)
4 MOhm // 1,2 pF
Input Impedance
(each side to ground)
2 MOhm // 2,3 pF
Output Voltage Swing
±8 V using 1 MΩ oscilloscope input
Offset (typical)
±5 mV
Noise (typical)
2 mVrms
Source Impedance
50 Ohm
Power Supply
9 V power adapter (included)
Weight
250 g
Dimensions
195 x 55 x 30 mm
BNC cable length
100 cm
Input lead length
55 cm each
Operation Tempearure/Humidity
0°C to 50°C / 10% to 85% RH
Storage Tempearure/Humidity
-30°C to 70°C / 10% to 90% RH
Safety Standard
IEC 61010-031:2015 CAT III
Pollution Degree
2
Specifications are subject to change without notice. To ensure the most current
version of this manual, please download the current version from our website: