1.8 Low Frequency CMRR Adjustment 

a. Referring to Fig. 9, connect output BNC to the oscilloscope which has good grounding. 
b. Connect both input lines to a 220/200VAC, 60/50Hz power source. 
c. Put SW1 at 1/10 (or 1/20). 
d. Set the VOLTS/DIV switch on oscilloscope’s panel to 1mV and the TIME/DIV switch to 5ms. Then 

a 60/50Hz sinusoid waveform should be displayed on the screen. 

e. Adjust VR1 to make the amplitude of displayed waveform as small as possible. Finally the 

amplitude should be less than 1.5mV. 

 

Fig. 9 

 
 
 
 
 
 
 
 

 

5

Summary of Contents for SI-9001

Page 1: ...utes to keep it warm before the calibration starts Then follow the steps below Offset Adjustment A Referring to the Figure below connect both input lines black and red together thus actually shorting...

Page 2: ...owever when VR4 is changed the Offset at 1 10 is affected thus we must calibrate again the 1 10 by repeating steps C and D G Again adjust the Offset at 1 100 by repeating steps E and F Through this co...

Page 3: ...1 1 Remove the battery lid 1 2 You will see the two screws and loosen them Fig 1 1 3 Near the BNC cable side there are two protuberant mortises Use fingers to push back and pull upward Fig 2 1 4 Use f...

Page 4: ...on Fig 4 1 6 You will see the two metal cases There is one screw on the PCB Loosen it then the whole metal case can be taken out There are two wires still connected with the upper case Leaves them the...

Page 5: ...ase connect a 6VDC power adaptor to CON1 as shown in the Figure above and turn on SW2 the LED indicator must light up Keep the unit turned on for 10 minutes to keep it warm before the calibration star...

Page 6: ...be F Again referring to the Figure above adjust VR4 to make output voltage approach to zero This way the Offset is adjusted into range within 0 1mV However when VR4 is changed the Offset at 1 10 is af...

Page 7: ...a 220 200VAC 60 50Hz power source c Put SW1 at 1 10 or 1 20 d Set the VOLTS DIV switch on oscilloscope s panel to 1mV and the TIME DIV switch to 5ms Then a 60 50Hz sinusoid waveform should be displaye...

Page 8: ...ine to ground c Put SW1 at 1 10 or 1 20 d Set the VOLTS DIV switch on oscilloscope s panel to o 1V for 1 10 or 1 20 attenuation and the TIME DIV switch to 20us e Adjust VC1 to make the displayed wavef...

Page 9: ...the oscilloscope has good grounding b Connect both input lines to a 155Vp p 20KHz sine wave signal source c Put SW1 at 1 10 1 20 d Set VOLTS DIV switch on oscilloscope s panel to 1mV and the TIME DIV...

Page 10: ...ent a Referring to Fig 13 connect output BNC to a DMM b Connect the red input line to the positive terminal of a DC voltage source 70VDC and the black input line to the negative terminal source 70VDC...

Page 11: ...d Adjust VR3 to make the output voltage equal to 6 9V or 3 51V for 1 20 Fig 13 9...