AC current measurement / A~
Set the appropriate range using the selector switch. Connect the black test lead to the
“COM” jack and the red test lead to the mA or 10 A jack. Connect test leads to the test
specimen. Read off measurement result from the display.
You must use the “10 A” jack when measuring currents above
500 mA!
Alternating current
Measuring
range
Resolution
Accuracy
40 Hz
– 1 kHz 1 kHz – 10 kHz 10 kHz – 20 kHz
500 µA
0.01 µA
± 0.75 % of
meas. val.
+ 20 digits
± 1 % of
meas. val.
+ 20 digits
± 1.2 % of
meas. val.
+ 5 digits
5000 µA
0.1 µA
± 0.75 % of
meas. val.
+ 10 digits
± 1 % of
meas. val.
+ 10 digits
± 1.2 % of
meas. val.
+ 5 digits
50 mA
1 µA
± 0.75 % of
meas. val.
+ 20 digits
± 1 % of
meas. val.
+ 20 digits
± 1.2 % of
meas. val.
+ 5 digits
500 mA
10 µA
± 0.75 % of
meas. val.
+ 10 digits
± 1 % of
meas. val.
+ 10 digits
± 1.2 % of
meas. val.
+ 5 digits
5 A
0.1 mA
± 0.75 % of
meas. val.
+ 20 digits
± 1.5 % of
meas. val.
+ 20 digits
± 1.2 % of
meas. val.
+ 5 digits
10 A*
1 mA
± 1 % of
meas. val.
+ 10 digits
± 1.5 % of
meas. val.
+ 10 digits
not specified
Accuracy specified only valid for 10 % - 100 % of the measuring range.
- 500 mA range is protected by a F 500 mA / 1000 V fuse.
- 10 A range is protected by a F 10 A / 1000 V fuse.
- in the 10 A range observe the maximum duty cycles!
- Frequency range: 40
– 20000 Hz.
* Following measurement, which should take a
max. 10 seconds, allow the instrument to cool down
for 15 minutes to protect against overheating.
Press the SELECT button to switch between DC, AC and AC+DC measurement
methods.
Summary of Contents for TB 312
Page 1: ...Testboy TB 312 Version 1 5...
Page 196: ...196 Testboy TB 312 CE...
Page 197: ...Testboy TB 312 197 1 2 3 2 4 5 CE Testboy GmbH...
Page 198: ...198 Testboy TB 312...
Page 199: ...Testboy TB 312 199 Testboy Testboy Testboy Testboy WEEE...
Page 200: ...200 Testboy TB 312 Cd Hg Pb Testboy GmbH Testboy GmbH www testboy de...
Page 201: ...Testboy TB 312 201 Testboy TB 312 TB 312 CAT III 1000 USB TB 312 CAT III 1000...
Page 202: ...202 Testboy TB 312 70 35 DC 33 16 AC DIN VDE EN 61010 1 CAT II CAT III CAT IV...
Page 205: ...Testboy TB 312 205 8 9 10 AC 11 12 DC 13 14 Please Wait 50 F 5000 F 15 16 PCLINK USB 17...
Page 211: ...Testboy TB 312 211 COM V mV 0 1 mV 0 7 mA 2 5 V SELECT COM V mV 60 60...
Page 213: ...Testboy TB 312 213 6 x 1 5 V AAA Micro F 10 A 1000 V F 500 mA 1000 V...
Page 214: ...214 Testboy TB 312 Setup exe...
Page 215: ...Testboy TB 312 215 Next Next Install Finish Programme Multimeter Multimeter exe Multimeter exe...
Page 217: ...Testboy TB 312 217 XY X Y Druck...