background image

 

8/8 

6

 

Basic Maintenance 

 
Accumulator charging: 

   During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if 
necessary. If accumulator is discharged the probe automatically turns off. The probe is switched on again 
powering USB port.   Current state of accumulator can be determined in three ways:

 

1) 

By LED diode ´STAT´

 - If LED starts blinking in green-yellow color it indicates that system is working 

incorrectly and one of main case is low voltage of accumulator. (see paragraph 

Operation Manual / 

LED diode “STAT””

)  

2)  On  wireless  Central  Unit  display 

  Symbol  "#"  on  left  edge  of  display  second  line  means  that 

accumulator voltage is low. 
3) In TERAview and usbRADONview application - where you can check current accumulator voltage. 
Voltage should not fall below 3.5 V, in limit conditions falls below 3.3V. 

Accumulator  is  charged  via  USB  port  using  supplied  USB  cable.  USB  cable  can  be  connected  to  PC  or  to 
supplied power adapter.  
 

Dehumidifier replacing 

  For correct measuring and low consumption the probe has to work in environment with relative humidity 
max  up  to  90%.  In  places  with  higher  humidity  it  is  needed  to  insert  1  dehumidifier  into  2  chambers  of 
probes. For inserting dehumidifier to probe is necessary to screw and disassemble 2. and 3. chamber from 
the bottom; see picture in chapter2. If the relative humidity of environment is above 90% and temperature 

10 °C the change interval of dehumidifier is 

6 weeks. Humid dehumidifier is possible to use repeatedly after 

drying it up in oven with 210

°C for 

1 hour. 

The temperature cannot be  over 230°C.

  

7

 

EC Declaration of Conformity 

EC  Declaration  of  Conformity  will  be  delivered  by  Tesla  producer  on  request.  If  interested,  please  use 
contacts on the web 

www.tesla.cz

 

8

 

Repairs 

 

Any repairs and non basic maintenance must be performed exclusively by TESLA manufacturer. 
 

TESLA 

Podebradska 56/186 

180 66 Prague 9 

www.tesla.cz

  

9

 

Warranty 

-

 

This product is covered by warranty of 24 months from purchase date. 

-

 

In case of warranty claim, please contact our Service Department. 

-

 

Warranty covers any defects in materials or workmanship and excludes any damage resulting from 

or caused by transport or handling or by any misuse. 

-

 

Warranty ceases if product has been used improperly or its seal is broken. 

-

 

In case of warranty claim, warranty period is prolonged by number of days product was undergoing 

warranty repairs.  

-

 

After the end of its life, product must be handled as e-waste.

 

10

 

Accessories 

Radon Probe accessories are available at produce

www.tesla.cz

 or at distributor. 

Reserve antenna            

 

 

                                 

11

 

Revision History 

    

Revision 

Date 

Comments 

Rev.1:

 

31.1.2017 

Initial release 

 

 

Summary of Contents for TSRP3

Page 1: ...chnical Specifications Operation Manual v 1 2019 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 6 4 Product Specification 6 5 Operating Instructions 7 6 Basic Maintenance 8 7 EC Declaration of Conformity 8 8 Repairs 8 9 Warranty 8 10 Accessories 8 11 Revision History 8 ...

Page 2: ...d product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specifications and user operation of the TSRP3 WLAN Wireless and USB Radon Soil Probe Product was developed and manufactured in the Czech Republic All rights reserved TESLA Offer or delivery of products or services re...

Page 3: ...dict seismic activity or earthquakes Portable probe basis is a measuring chamber with a semiconductor photodetector Radon enters the chamber by diffusion through the input filter on the bottom of probe The probe measures in autonomous and time continuous way It processes results every 2 minute intervals and from this counts short term moving average of radon concentration 1 hour moving average ave...

Page 4: ...ntral unit memory Via connected computer to Central Unit and TERAview application on PC it is possible to download and process all data from system and configure whole system Setting and configuration of Radon Soil Probe and whole system is also managed by TERAview application on PC TERAview application version 3 11 6 and higher drivers and user manual with detail configuration description is free...

Page 5: ...ion and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuator in central unit can be use which is hardwired with power relay Power relay switches on a fan which decreases radon concentration within an area After decreasing of radon concentration actuator receives command to switch off fan This cycle repeats depending on increasing or decr...

Page 6: ...m RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring temperature range 20 to 60 C Radio interface 868MHz Max number of measuring network elements 16 Probe to terminal unit distance RF range depends on number of walls and building material up to 600 m in open space Possibility of using repeater for RF range extending yes Result...

Page 7: ...livered power adapter Power adapter is connected to probe via provided USB cable In case of blackout internal accumulator ensures UPS function Configuration Setting and configuration is different according to operation way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application...

Page 8: ...fier to probe is necessary to screw and disassemble 2 and 3 chamber from the bottom see picture in chapter2 If the relative humidity of environment is above 90 and temperature 10 C the change interval of dehumidifier is 6 weeks Humid dehumidifier is possible to use repeatedly after drying it up in oven with 210 C for 1 hour The temperature cannot be over 230 C 7 EC Declaration of Conformity EC Dec...

Reviews: