background image

 

4/8 

Radon Probe can be operated by these ways: 

 

A)

 

Standalone  probe

  -  Thanks  to  its  independent  accumulator  power,  portable  radon  measuring 

probe supports flexible placing options within monitored structures. Accumulator will last for more 
than 1 year after full charging. The resulting values are downloaded after end of the measurement 
by B) or C) way. 

 

 

B)

 

Probe  connected  via  USB

 

  Using  usbRADONview  app  and  USB  cable  is  possible  to  download 

results  to  PC  continuously  during  measurement  or  at  once  at  the  end  of  measurement. 
usbRADONview application, drivers and user manual is free downloaded on website of producer. 

 

 

C1)   

Wireless network for radon measurement

 - Central Unit supports simultaneous data downloads 

from  up  to  16  probes.  All  values  are  saved  again  into  central  unit  memory.  Via  connected 
computer to Central Unit and TERAview application on PC it is possible to download and process 
all data from system and configure whole system.  Setting and configuration of Radon Soil Probe 
and whole system is also managed by TERAview application on PC. TERAview application version 
3.11.6  and  higher,  drivers  and  user  manual  with  detail  configuration  description  is  free 
downloaded on website of producer.  

   In  case  of  time  continuous  measurement  of  radon  concentration  or  in  case  of  setting  in 

regulation system the probe must be placed in radio range of Central Unit. Distance (radio range) 
between probes and central unit is up to 600m in open space. In buildings it depends on number 
of  walls,  building  material,  etc.  Strength  of  radio  signals  (RSSI)  is  monitored  by  Central  Unit.  If 
radio  signal  strength  between  individual  elements  is  insufficient,  another  radon  probe  or 
repeater TRR must be inserted and used to extend the signal.

 

 

  

Summary of Contents for TSRP3

Page 1: ...chnical Specifications Operation Manual v 1 2019 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 6 4 Product Specification 6 5 Operating Instructions 7 6 Basic Maintenance 8 7 EC Declaration of Conformity 8 8 Repairs 8 9 Warranty 8 10 Accessories 8 11 Revision History 8 ...

Page 2: ...d product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specifications and user operation of the TSRP3 WLAN Wireless and USB Radon Soil Probe Product was developed and manufactured in the Czech Republic All rights reserved TESLA Offer or delivery of products or services re...

Page 3: ...dict seismic activity or earthquakes Portable probe basis is a measuring chamber with a semiconductor photodetector Radon enters the chamber by diffusion through the input filter on the bottom of probe The probe measures in autonomous and time continuous way It processes results every 2 minute intervals and from this counts short term moving average of radon concentration 1 hour moving average ave...

Page 4: ...ntral unit memory Via connected computer to Central Unit and TERAview application on PC it is possible to download and process all data from system and configure whole system Setting and configuration of Radon Soil Probe and whole system is also managed by TERAview application on PC TERAview application version 3 11 6 and higher drivers and user manual with detail configuration description is free...

Page 5: ...ion and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuator in central unit can be use which is hardwired with power relay Power relay switches on a fan which decreases radon concentration within an area After decreasing of radon concentration actuator receives command to switch off fan This cycle repeats depending on increasing or decr...

Page 6: ...m RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring temperature range 20 to 60 C Radio interface 868MHz Max number of measuring network elements 16 Probe to terminal unit distance RF range depends on number of walls and building material up to 600 m in open space Possibility of using repeater for RF range extending yes Result...

Page 7: ...livered power adapter Power adapter is connected to probe via provided USB cable In case of blackout internal accumulator ensures UPS function Configuration Setting and configuration is different according to operation way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application...

Page 8: ...fier to probe is necessary to screw and disassemble 2 and 3 chamber from the bottom see picture in chapter2 If the relative humidity of environment is above 90 and temperature 10 C the change interval of dehumidifier is 6 weeks Humid dehumidifier is possible to use repeatedly after drying it up in oven with 210 C for 1 hour The temperature cannot be over 230 C 7 EC Declaration of Conformity EC Dec...