RS-232C Command List
37
5.5.1 Error Data List
Table 5-3
List of Error Data
Error Code
Description
ERR10
Syntax error. The command does not exist or is invalid.
ERR20
Handler inward sensor error. It only occurs when the Handler is in Sensor Mode.
(The horizontal operation of fixture is incomplete.)
ERR30
Probe downward sensor error. It only occurs when the Probe is in Sensor Mode.
(The vertical operation of fixture is incomplete.)
ERR40
Status error, It occurs when an invalid command is sent. For example, a Handler
OUTWARD or Probe UPWARD while the lid is closed. It is returned if the command
is sent when the motion can not be done in conjunction with the current status.
ERR50
Lid open sensor error. It occurs only to certain products which should always detect
whether the lid is open. The response comes when the lid open sensor does not
detect the cylinder in open position.
ERR60
Safety error. The Lid Close Sensor does not detect the cylinder in close position.
ERR90
It occurs when the command is sent while the Handler or the Probe is in Off Mode.
Summary of Contents for TC-5972DP
Page 1: ...TC 5972DP Pneumatic Shield Box User Manual R20191111...
Page 9: ...PART 1 1 Getting Started...
Page 13: ...PART 2 5 Hardware Description...
Page 14: ...2 Hardware Overview Front 6 2 1 Front Figure 2 1 TC 5972DP Front Components...
Page 20: ...12 Dimensions 2 5 2 Outer Dimensions...
Page 26: ...18 Fixtures Figure 3 1 Sample Images of Custom Fixtures...
Page 27: ...PART 3 19 Putting into Operation...
Page 54: ...PART 4 46 Maintenance...
Page 62: ...54 Expendables and Accessories Figure 7 1 TC 5972DP Gasket Locations...