![Tescom TC-5915AP User Manual Download Page 52](http://html1.mh-extra.com/html/tescom/tc-5915ap/tc-5915ap_user-manual_1090866052.webp)
44
RS-232C Commands in Detail
5.6.6 PASS/FAIL Commands (While The Lid is Closed)
PASS
A command that decides PASS after a test.
If the PASS command is sent, the PASS LED on the control box will glow green (
), and the lid
will open automatically.
Type
Command
Input
PASS
Response
{ OK | ERR40 }
●
OK : turns the green PASS LED on, and the lid opens.
●
ERR40: Status error, The command is sent while the lid is open.
FAIL
A command that decides FAIL after a test.
The FAIL LED on the control box will glow red (
), and the lid will
NOT
open.
Type
Command
Input
FAIL
Response
{ OK | ERR40 }
●
OK : turns the red FAIL LED on.
●
ERR40: Status error. The command is sent while the lid is open.