40
RS-232C Commands in Detail
CLOSE
A command to close the lid.
The lid-close sensor on a cylinder will blink red if the lid is closed regularly.
Type
Command
Input
CLOSE
Response
{ READY | ERR20 | ERR30 | ERR60 }
●
READY : Lid closed. Ready to test.
●
ERR20 : Handler-inward sensor error. It occurs only when the Handler is in Sensor
Mode. (The horizontal operation of the fixture is incomplete.)
●
ERR30 : Probe-downward sensor error. It occurs only when the Probe is in Sensor
Mode. (The vertical operation of the fixture is incomplete.)
●
ERR60 : Safety error. The lid-close sensor does not detect the lid..
LID?
A query to check the current status of the lid.
Type
Query
Input
LID?
Response
{ OPEN | CLOSE }
●
OPEN : Lid opened
●
CLOSE : Lid closed
5.6.4 Handler Operation Commands
HANDLER?
INWARD
OUTWARD