KATC-C1 Instructions
Page 26
04/14
TEST CURRENT OUTPUT
This section describes the test current output, which is available by sending the KATC-C1 the test
current command from a computer. This command is described in a preceding section. The purpose of
the command is to allow use of the KATC-C1 in an automated environment. In such an environment
it may be useful to know the present value of the transformer-under-test secondary current as an aid in
automatically adjusting the test current to the desired value.
ASCII format:
{CR}{LF}
XX.XX{CR}{LF}
{CR}{LF}
Record layout:
1
+---+
¦*¦#¦
+---+
3
+-------------+
¦X¦X¦.¦X¦X¦*¦#¦
+-------------+
10
+---+
¦*¦#¦
+---+
Notes: When the TUT Test Current Switch is set to 1 ampere, the decimal point is shifted one digit to
the left (X.XXX).