TESA-hite Magna 400 / 700
35
4
44
4....14
14
14
14 U
U
U
Use of the fine adjustment
se of the fine adjustment
se of the fine adjustment
se of the fine adjustment
The fine adjustment is used when one wants to precisely adjust a height in 3 «ZZ» mode. It can also
be used when seeking the culmination point of a bore or a shaft. Indeed, in such a case, the probing
force can be very finely adjusted affording great stability.
10
Lightly block
the light grey
light grey
light grey
light grey
knurled screw
9
Pull and turn
the dark grey
dark grey
dark grey
dark grey
knurled screw
5 INSTRUMENT CONFIGURATION
To enter into the configuration mode, you have to activate by turned on the instrument, first the F1,
then simultaneously the F1 and ON/OFF keys; after deactivate the ON/OFF and then the F1 key.
The configuration menu allows the following functions:
-
F1: Activates or deactivates the Beep, when probing or detecting the culmination point.
By default the Beep strong is active (BEEP Hi).
-
F2: By default the TESA-Hite Magna 400 / 700 turns off after 20 min (AUTO OF).
To turn it off only with the ON/OFF button, set this option on ON.
-
F3: Change the resolution of the instrument (0.001 / 0.005 / 0.01 mm).
By default the resolution is 0.001 mm.
Summary of Contents for TESA-hite Magna 400
Page 2: ...TESA hite Magna 400 700 2 ...