Quickstart document
20
Mode Description
ST1
Lengths measurement in only one probing direction.
The calibration of the probe is not necessary.
ST2
Lengths measurement in two probing directions.
The calibration of the probe is mandatory.
Measurement with one probing direction
Measurement with two probing directions
ST1
The access to
ST1
mode does not require the determination of the probe constant. After
entering in this mode, the first mandatory step is to define a reference.
All measurements regarding the same reference have to be carried out by probing in a
direction similar to the one chosen when the active reference has been captured.
ST2
Each time when entering
ST2
mode, it is necessary to calibrate the probe
. This must
be done using the masterpiece delivered with your instrument. For more details regarding
this process please refer to the user manual. Once this has been carried out, the software
will ask you to define a reference (always by double probing).
Then you can perform your measurements in both probing directions (up or down).