1
2.
SPECIFICATIONS
2-1 Electrical Specifications
Measurement Range :
Type J :
-200
℃
to 760
℃
-328
℉
to 1400
℉
Type K :
-200
℃
to 1370
℃
-328
℉
to 2498
℉
Resolution : 0.1
℃
, 1
℃
, 0.2
℉
, 2
℉
Maximum Voltage at Thermocouple Input :
60V dc ,or 24Vrms ac
RF Field Derating :
Strong RF fields and low-frequency adversely affect accurate
measurement .
Operating Temperature and Humidity:
0
℃
to 40
℃
(32
℉
to 104
℉
), 0 - 80
%
RH
Storage Temperature and Humidity:
-10
℃
to 60
℃
(14
℉
to 140
℉
), 0 - 70
%
RH
2
Basic Accuracy:
( @23
±
5
℃
Calibration ) Accuracy are
±
( ...
%
of
r degree ) at 18
℃
to 28
℃
with relative humidity up to 80
%
For single thermocouple measurements
Function Resolution
Range
Accuracy
Type K
Type J
℃
0.1
℃
-200
℃
~-50
℃
-200
℃
~-50
℃
(0.5%rdg+1.0
℃
)
-50
℃
~0
℃
-50
℃
~0
℃
(0.2%rdg+1.0
℃
)
0
℃
~200
℃
0
℃
~150
℃
(0.1%rdg+ 0.8
℃
)
1
℃
200
℃
~1370
℃
150
℃
~760
℃
(0.2%rdg+ 2
℃
)
℉
0.2
℉
-328
℉
~-58
℉
-328
℉
~-58
℉
(0.5%rdg+ 2.0
℉
)
-58
℉
~32
℉
-58
℉
~32
℉
(0.2%rdg+ 2.0
℉
)
32
℉
~392
℉
32
℉
~302
℉
(0.1%rdg+1.6
℉
)
2
℉
392
℉
~2498
℉
302
℉
~1400
℉
(0.2%rdg+3
℉
)
For T1-T2 Measurements accuracy is Basic accuracy add 0.2%rdg.
NOTE
The basic accuracy specification does not include the error of the
probe. Please refer to the probe accuracy specification for
additional details.
Temperature Coefficient:
For ambient temperatures from 0
℃
to 18
℃
and 28
℃
to 40
℃
Summary of Contents for TES-1306