SPECIFICATIONS 

2-1 Electrical Specifications 

Measurement Range : 

Type J :  
-200

 to 760

 

-328

 to 1400

 

Type K :  
-200

 to 1370

 

-328

 to 2498

 

Resolution : 0.1

 , 1

 , 0.2

 , 2

 

Maximum Voltage at Thermocouple Input :  

60V dc ,or 24Vrms ac 

RF Field Derating : 

Strong RF fields and low-frequency adversely affect  accurate  
measurement . 

Operating Temperature and Humidity: 

0

 to 40

 (32

 to 104

 ),  0 - 80

RH 

Storage Temperature and Humidity: 

-10

 to 60

 (14

 to 140

 ),  0 - 70

RH 

 
 
 
 
 

 
 

Basic Accuracy:

 ( @23

±

 5

 Calibration ) Accuracy  are

±

 ( ...

 of 

r degree ) at 18

 to 28

 with relative humidity up to 80

 

For single thermocouple measurements 

Function  Resolution 

Range 

Accuracy 

Type K 

Type J 

 

0.1

 

-200

~-50

  -200

~-50

 

(0.5%rdg+1.0

-50

~0

 

-50

~0

 

(0.2%rdg+1.0

0

~200

 

0

~150

 

(0.1%rdg+ 0.8

1

 

200

~1370

  150

~760

 

(0.2%rdg+ 2

 

0.2

 

-328

~-58

  -328

~-58

 

(0.5%rdg+ 2.0

-58

~32

 

-58

~32

 

(0.2%rdg+ 2.0

32

~392

  32

~302

 

(0.1%rdg+1.6

2

 

392

~2498

  302

~1400

 

(0.2%rdg+3

For T1-T2 Measurements accuracy is Basic accuracy add 0.2%rdg. 

NOTE

 

The basic accuracy specification does not include the error of the 
probe.    Please  refer  to  the  probe  accuracy  specification  for 
additional details. 

Temperature Coefficient: 

For ambient temperatures from 0

 to 18

 and 28

 to 40

 

Summary of Contents for TES-1306

Page 1: ... or service the meter Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired Environment conditions Altitude up to 2000 meters Relatively humidity 90 max Operation Ambient 0 40 Maintenance Clearing Repairs or servicing not covered in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not ...

Page 2: ...uracy are of reading degree at 18 to 28 with relative humidity up to 80 For single thermocouple measurements Function Resolution Range Accuracy Type K Type J 0 1 200 50 200 50 0 5 rdg 1 0 50 0 50 0 0 2 rdg 1 0 0 200 0 150 0 1 rdg 0 8 1 200 1370 150 760 0 2 rdg 2 0 2 328 58 328 58 0 5 rdg 2 0 58 32 58 32 0 2 rdg 2 0 32 392 32 302 0 1 rdg 1 6 2 392 2498 302 1400 0 2 rdg 3 For T1 T2 Measurements accu...

Page 3: ...he display Break or no input indication _ _ _ _ appears on the display Low Battery Indication The is displayed when the battery voltage drops below the operating voltage Power Requirement 9 Volt battery NEDA 1604 or JIS 006P or IEC6F22 Battery Life typical 100hours Alkaline Battery Dimensions 135 L x 72 W x 31 H mm 5 3 L x 2 8 W x 1 2 H inches Weight Approx 235g with battery Accessories Battery In...

Page 4: ...in the meter will show T1 T2 T1 T2 in circulation If this key been pressed again it will turn the meter to default reading that will be T1 7 REL Press this key will turn the meter into relative mode The reading will first show zero and then the reading change relate to the setting time will be shown Press it again will show relative value If hold this key 2 seconds the meter will be back to the no...

Page 5: ...ion probe 50 to 1000 58 to 1832 2 2 or 0 75 3 6 or 0 75 3 2φ 150 metal sheath 100 Compensating wire TP K03 Surface probe 50 to 750 58 to 1382 2 2 or 0 75 3 6 or 0 75 100 Compensating wire 12 5φ x 94 handle TP K01 Available for general condition especially for complex and any place hard to reach 8 TP K02 Available for temperature measurement of liquid gels or air TP K03 Available for flat or curved...

Page 6: ...9 ...

Page 7: ...TES ELECTRICAL ELECTRONIC CORP 7F No 31 Lane 513 Rui Guang Road Neihu Dist Taipei Taiwan R O C Tel 02 2799 3660 Fax 886 2 2799 5099 E Mail tes ms9 hinet net http www tes com tw June 2000 4 ...

Reviews: