Terumo Azur CX 35 Instructions For Use Manual Download Page 4

- 4 -

attached to the delivery pusher and attempt another detachment 
cycle when the light turns green.  

38. The light will turn red after the number of detachment cycles

specified on the Azur Detachment Controller labeling.  DO NOT
use the Azur Detachment Controller if the light is red.  Discard the
Azur Detachment Controller and replace it with a new one when
the light is red.

39. Verify detachment of the coil by first loosening the RHV valve,

then pulling back slowly on the delivery system and verifying that
there is no coil movement.  If the implant did not detach, do not
attempt to detach it more than two additional times.  If it does not
detach after the third attempt, remove the delivery system.

40. After detachment has been confirmed, slowly retract and remove

the delivery pusher. 

Advancing the delivery pusher once the

coil has been detached involves the risk of aneurysm or
vessel rupture. Do NOT advance the delivery pusher once the
coil has been detached.

41. Verify the position of the coil angiographically.
42. Additional coils may be deployed into the lesion as described

above.  Prior to removing the catheter from the treatment site,
place an appropriately sized guidewire completely through the
catheter lumen to ensure that no part of the last coil remains
within the catheter.

The physician has the discretion to modify the coil deployment 
technique to accommodate the complexity and variation in 
embolization procedures.  Any technique modifications must be 
consistent with the previously described procedures, warnings, 
precautions and patient safety information. 

SPECIFICATIONS FOR AZUR DETACHMENT CONTROLLER 

Output voltage: 8 ± 1 VDC

Cleaning, preventative inspection, and maintenance: The
Azur Detachment Controller is a single use device,
preloaded with battery power, and packaged sterile.  No
cleaning, inspection, or maintenance is required.  If the
device does not perform as described in the Detachment
section of these Instructions, discard the Azur Detachment
Controller and replace it with a new unit.

The Azur Detachment Controller is a single use device.  Do
not reuse, reprocess or resterilize. Reuse, reprocessing or
resterilization may compromise the structural integrity of the
device and/or lead to device failure which, in turn, may result
in patient injury, illness, or death. Reuse, reprocessing, or
resterilization may also create a risk of contamination of the
device and/or cause patient infection or cross-infection,
including, but not limited to, the transmission of infectious
disease(s) from one patient to another. Contamination of the
device may lead to injury, illness or death of the patient.

Batteries are pre-loaded into the Azur Detachment
Controllers.  Do not attempt to remove or replace the
batteries prior to use.

After use, dispose of the Azur Detachment Controller in a
manner consistent with local regulations.

PACKAGING AND STORAGE 

The Azur system is placed inside a protective, plastic dispenser 
hoop and packaged in a pouch and unit carton.  The Azur system 
and dispenser hoop will remain sterile unless the package is 
opened, damaged, or the expiration date has passed.  Store at a 
controlled room temperature in a dry place. 

The Azur Detachment Controller is packaged separately in a 
protective pouch and carton.  The Azur Detachment Controller 
has been sterilized; it will remain sterile unless the pouch is 
opened, damaged, or the expiration date has passed. Store at a 
controlled room temperature in a dry place. 

SHELF LIFE 

See the product label for the device shelf life.  Do not use the 
device beyond the labeled shelf life. 

MR INFORMATION

 

Non-clinical testing demonstrated that the coil implant is

 MR 

conditional

. A patient can be scanned safely, immediately after 

placement under the following conditions: 

Static magnetic field of 1.5 and 3 Tesla, with

Maximum spatial gradient field of 5000 Gauss/cm

Theoretically estimated maximum whole body average
(WBA) specific absorption rate (SAR) of < 2W/kg
(Normal Operating Mode)

MRI-Related Heating 
Under the scan conditions above, and 15 minutes of continuous 
scanning, the coil implant is expected to produce a maximum 
temperature rise of 1.3°C  

Image Artifact Information 

MR image quality may be compromised if the area of interest is in 
the exact same area or relatively close to the position of the coil 
implant. Therefore, optimization of MR imaging parameters to 
compensate for the presence of this device may be necessary. 

In non-clinical testing, the image artifact caused by the coil 
implant extends approximately 41.3mm in diameter and 21.3mm 
in height from the implant when imaged with a gradient and spin 
echo pulse sequence and a 3 Tesla MR system. 

Terumo Corporation recommends that the patient register the MR 
conditions disclosed in this IFU with the MedicAlert Foundation or 
equivalent organization. 

MATERIALS 

The Azur system does not contain latex or PVC materials. 

SYMBOLS 

Lot Number 

Order Number 

Content 

Sterilized Using Irradiation 

Sterilized Using Ethylene Oxide 

Do Not Reuse 

Use-by Date 

Date of Manufacture 

Attention, Consult Accompanying Documents 

Type BF Applied Part 

Summary of Contents for Azur CX 35

Page 1: ...gle use only Do not reuse reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in...

Page 2: ...timate the size of the lesion to be treated 9 For aneurysm occlusion the diameter of the first and second coils should never be less than the width of the aneurysm neck or the propensity for the coils...

Page 3: ...otating the delivery pusher may result in a stretched coil or premature detachment of the coil from the delivery pusher which could result in coil migration Angiographic assessment should also be perf...

Page 4: ...ction including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may lead to injury illness or death of the patient Batteries are pre...

Page 5: ...age or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability or responsibility in...

Page 6: ...use reprocess or resterilize Reuse reprocessing or resterilization may compromise the structural integrity of the device and or lead to device failure which in turn may result in patient injury illnes...

Page 7: ...will allow for fluoroscopic road mapping during the procedure 6 Select a microcatheter with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guid...

Page 8: ...the introducer sheath Excessive tightening could damage the device 23 Push the coil into the lumen of the microcatheter Use caution to avoid catching the coil on the junction between the introducer s...

Page 9: ...e risk of aneurysm or vessel rupture Do NOT advance the delivery pusher once the coil has been detached 43 Verify the position of the coil angiographically through the guide catheter 44 Additional coi...

Page 10: ...not limited to any implied warranties of merchantability or fitness for particular purpose Handling storage cleaning and sterilization of the device as well as factors relating to the patient diagnos...

Page 11: ...ation of the device and or cause patient infection or cross infection including but not limited to the transmission of infectious disease s from one patient to another Contamination of the device may...

Page 12: ...e may be increased For vessel occlusion it is suggested that the diameter of the initial coil placed be slightly larger than the actual vessel diameter to prevent displacement or migration For aneurys...

Page 13: ...catheter tip from moving away from the intended delivery location 20 Once the coil has been deployed and prior to removing the delivery catheter advance the guidewire through the catheter tip to verif...

Page 14: ...sequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for it any other or additional liability o...

Page 15: ...device and or lead to device failure which in turn may result in patient injury illness or death Reuse reprocessing or resterilization may also create a risk of contamination of the device and or caus...

Page 16: ...er with the appropriate inner diameter After the microcatheter has been positioned inside the lesion remove the guidewire COIL SIZE SELECTION 7 Perform fluoroscopic road mapping 8 Measure and estimate...

Page 17: ...ppropriately sized coil Movement of the coil may indicate that the coil could migrate once it is detached DO NOT rotate the delivery pusher during or after delivery of the coil into the vasculature Ro...

Page 18: ...of the patient Batteries are pre loaded into the Azur Detachment Controllers Do not attempt to remove or replace the batteries prior to use After use dispose of the Azur Detachment Controller in a man...

Page 19: ...not be liable for any incidental or consequential loss damage or expense directly or indirectly arising from the use of this device Terumo neither assumes nor authorizes any other person to assume for...

Page 20: ...PD110323 Rev A 1 MicroVention Inc Instructions for Use Detachable...

Page 21: ...sitioned in the vessel or aneurysm within the specified reposition time from the time the device is first introduced into the microcatheter If the coil cannot be positioned and detached within this ti...

Page 22: ...ation replace the device If the green light remains solid green for the entire three second observation continue using the device 15 Hold the device just distal to the shrink lock and pull the shrink...

Page 23: ...The coil implant has been determined to be MR conditional according to the terminology specified in the American Society for Testing and Materials ASTM International Designation F2503 08 Non clinical...

Reviews: