&RQILUPWKDWWKHUHIHUHQFHLQGLFDWRURQWKHLQVHUWLRQVKHDWKLVIDFLQJXS7RHQVXUHSURSHURULHQWDWLRQRIWKH$QJLR6HDOGHYLFHZLWKWKHVKHDWKWKHVKHDWKFDSDQGWKHGHYLFH
VOHHYHRQO\ILWWRJHWKHULQWKHFRUUHFWSRVLWLRQ7KHUHIHUHQFHLQGLFDWRURQWKHGHYLFHFDSVKRXOGDOLJQZLWKWKHUHIHUHQFHLQGLFDWRURQWKHLQVHUWLRQVKHDWKFDS)LJXUH.HHSLQJWKH
LQVHUWLRQVKHDWKLQSODFHFDUHIXOO\DGYDQFHWKH$QJLR6HDOGHYLFHLQVPDOOLQFUHPHQWVXQWLOFRPSOHWHO\LQVHUWHGLQWRWKHLQVHUWLRQVKHDWK7KHVKHDWKFDSDQGWKHGHYLFHVOHHYHZLOO
VQDSWRJHWKHUZKHQSURSHUO\ILWWHG
Fig. 8
NOTE: If significant resistance to carrier tube advancement is encountered when insertion is almost complete, the anchor may be impinging on the posterior wall of
the artery. DO NOT CONTINUE TO ATTEMPT TO ADVANCE. In this case, slight repositioning of the sheath, either by reducing the angle of the sheath with respect to
the skin surface or by pulling the sheath back by 1-2 mm, may permit normal deployment.
:LWKRQHKDQGFRQWLQXHWRKROGWKHLQVHUWLRQVKHDWKFDSVWHDG\WRSUHYHQWPRYHPHQWRIWKHVKHDWKLQWRRURXWRIWKHDUWHU\:LWKWKHRWKHUKDQGJUDVSWKHGHYLFHFDSDQGVORZO\
DQGFDUHIXOO\SXOOEDFN6OLJKWUHVLVWDQFHZLOOEHIHOWZKHQWKHGHYLFHFDSLVSXOOHGRXWIURPUHDUKROGLQJSRVLWLRQ&RQWLQXHSXOOLQJRQWKHGHYLFHFDSXQWLOUHVLVWDQFHIURPWKHDQFKRU
FDWFKLQJRQWKHGLVWDOWLSRIWKHLQVHUWLRQVKHDWKLVIHOW
7RHQVXUHWKHFRUUHFWDQFKRUSRVLWLRQFRQILUPWKDWWKHHGJHRIWKHGHYLFHFDSIDOOVZLWKLQWKHFRORUHGEDQGVRQWKHGHYLFHVOHHYH
Fig. 9 (Correct)
0DLQWDLQJULSRQWKHLQVHUWLRQVKHDWKDQGSXOOWKHGHYLFHKDQGOHVWUDLJKWEDFNLQWRWKHIXOOUHDUORFNHGSRVLWLRQ)LJXUH5HVLVWDQFHZLOOEHIHOWDVWKHGHYLFHKDQGOHDQGVOHHYH
VQDSORFN7KHGHYLFHVOHHYHFRORUHGEDQGVVKRXOGQRZEHFRPSOHWHO\YLVLEOH
NOTE: If the device sleeve separates from the sheath while attempting full rear lock positioning, do not push the Angio-Seal™ device forward to reattach the sheath
cap. Complete sealing of the puncture following instructions in C-1.
&RORUHG%DQG
Fig. 10
,QFRUUHFW,QGLFDWRU$OLJQPHQW
The Distal end of device cap completely covers the colored indicator band
RQWKHGHYLFHVOHHYH)LJXUH,IWKHDQFKRUFDWFKHVSUHPDWXUHO\DVLQ)LJXUHDGYDQFHWKHGHYLFH
LQWRWKHLQVHUWLRQVKHDWKDJDLQ,WPD\EHQHFHVVDU\WRSXVKWKHGHYLFHFDSEDFNLQWRWKHUHDUKROGLQJSRVLWLRQLQRUGHUWRJHWIXOOH[WHQVLRQRIWKHDQFKRUIURPWKHVKHDWK7KHQ
ZLWKGUDZWKHGHYLFHXQWLOWKHDQFKRUFDWFKHVFRUUHFWO\
Fig. 11 (Incorrect)
NOTE: Do not proceed until you are certain that the anchor has been properly deployed (Figure 9). If the anchor is improperly deployed, the Angio-Seal device will
not function.