©2012 Terumo BCT, Inc.
Page 5 of 14
PN 306620452
Replacing the Wafer Cassette (Cartridge)
Step
Action
1
Ensure that the device is not currently replacing a wafer.
2
Press the EJECT (Cassette/Cartridge Release) button.
The far edge of the empty cassette (cartridge) pops up.
3
Remove the empty cassette (cartridge).
4
Verify that the label on the new wafer cassette (cartridge) is facing up.
5
Slide the cavity on the front edge of the new cassette (cartridge) onto the
metal tab at the front edge of the wafer cassette (cartridge) compartment.
6
Push down the back edge of the new wafer cartridge (cassette) until it
snaps into place.
7
Press the RESET button to align the clamps and to advance an unused
wafer, if one is not already in place
Note:
While the wafer cassette (cartridge) is being replaced, one unused wafer
remains in the device. That wafer is automatically advanced to the welding area
when the RESET button is pressed, if a wafer is not already in place.
Placing the Tubing
Step
Action
1
Place the blood product containers gently on the bag support shelves; do
not drop them onto the shelves.
2
Place a length of fluid-filled or non-fluid-filled tubing into the rear clamp
slot.
Tubing must conform to the specifications listed in section 3,
“Specifications,” of the operating instructions.