©2012 Terumo BCT, Inc.
Page 13 of 14
PN 306620452
4
Lift out the holder that contains the fuse.
5
Remove the fuse and replace it with a 4A (250V) slow-blow fuse.
6
Place the fuse back into the holder and insert the holder back into the
TSCD II device.
7
Turn the holder cover one-half turn clockwise.
8
Repeat steps 3 through 7 for the second fuse holder.
Emptying and Cleaning the Wafer Disposal Box
Step
Action
Empty the wafer disposal box on a routine basis or as required. Remove the wafer
disposal box immediately if the LCD display indicates
DISPOSAL BOX FULL
. (Normal
operations cannot resume until the wafer disposal box has been emptied.)
Removing and emptying the wafer disposal box
1
Pull on the insert on the front of the box to loosen it from the TSCD II
device.
2
Slide the box forward to remove it from the device.
3
Discard the used wafers into an appropriate container.
Caution:
Wafers that have been used on tubing that contained blood
products should be considered biohazardous waste and should be
handled appropriately, according to your institution's SOPs.
Cleaning the wafer disposal box
1
Dampen a cloth with a mild detergent, isopropyl alcohol, or a broad-
spectrum disinfectant.
2
Clean the surface and the inside of the box with the dampened cloth.