EMC TABLES
Table 2:
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The T5460 T-SEAL MOBILE is intended for use in the electromagnetic environment specified below. The customer or
the user of the T5460 T-SEAL MOBILE should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
6kV
8kV
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/
output lines
2kV
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1kV line(s)
to line(s)
± 2kV line(s)
to earth
1kV
2kV
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply in-
put lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 sec
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level
.