Condition Media
Note:
Use the information in “Performing Tasks” to complete this task.
The purpose of this task is to allow the media to reach equilibrium with the provided gas supply before
loading the cells. This task includes two separate steps:
•
Step 1: provides rapid contact between the media and the gas supply by using a high EC circulation
rate.
•
Step 2: maintains the system in a proper state until the operator is ready to load the cells.
Table 8-1 describes the types of solution that are needed to attach to each line when performing
Condition Media. These solutions and corresponding volumes are based on the factory default settings
for this task.
Table 8-1: Solutions for Condition Media
Line
Solution
Attached to the
Line
Required Volume (estimate based on factory default settings)
Cell
None
N/A
Reagent
None
N/A
IC Media
None
N/A
Wash
None
N/A
EC Media
Media without
protein
0.1 L plus 6 mL/hour
Condition Media pathway:
Task
>
System Management
>
Condition Media
1.
Confirm the settings for step 1 shown in Table 8-2.
Table 8-2: Step 1 Settings for Condition Media
Setting
Factory Default
Setting Options
IC Inlet
None
N/A
IC Inlet Rate
0 mL/min
N/A
IC Circulation Rate
100 mL/min
N/A
EC Inlet
EC Media
Reagent, IC Media, EC Media, Wash
EC Inlet Rate
0.1 mL/min
N/A
EC Circulation Rate
250 mL/min
N/A
System Management Tasks
8-6
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...