![Terraillon TH1009N-C Instruction Manual Download Page 25](http://html.mh-extra.com/html/terraillon/th1009n-c/th1009n-c_instruction-manual_1089743025.webp)
25
25
CONNECTED NON-CONTACT THERMOMETER
EN
Applied part of type BF (The applied part is the probe tip)
Disposal in accordance with Directive 2012/19/EU (WEEE)
Complies with the European Medical Device Directive
(93/42/EEC) and amended by directive 2007/47/EC
requirements. Notified body TÜV Rheinland (CE0197)
The name and the address of the manufacturer
The name and the address of the Authorized EC-
representative in Europe
CH
The name and the address of the authorised representative
in Swiss
Caution
Refer to Instruction Manual
The first number 2: Protected against solid foreign objects
of 12,5 mm
Ф
and greater. The second number: Protected
against vertically falling water drops when enclosure titled
up to 15º.
Vertically falling drops shall have no harmful effects when
the enclosure is titled at any angle up to 15º, on either side
of the vertical.
Date of Manufacture
NORMALIZED SYMBOLS
DECLARATION OF CONFORMITY
Hereby, Terraillon SAS declares that Thermo Smart complies with the essential
health and safety conditions. This product meets the essential requirements
and other relevant provisions of the directive RED 2014/53/ UE. The complete
declaration of conformity is available on https:// terraillonhelp. zendesk.com/hc/
en-us (Declaration of Conformity section). E.I.R.P : 8,31 dBm.
Caution:
This thermometer has been thoroughly tested and inspected to assure proper
performance and operation!
Power supply:
DC 3V (2 x AAA battery)
Mode of operation:
Adjusted mode
Reference body site:
Core
Body measuring range:
32.0°C ~ 42.9°C (89.6°F ~ 109.3°F)
Laboratory accuracy:
±0.2°C(±0.4°F) for
35.0°C (95.0°F) ~ 42.0°C (107.6°F)
±0.3°C (±0.5°F) for other range
Object measuring range:
-22°C ~ 80.0°C (-7.6°F ~ 176.0°F)
Object measuring accuracy:
±2.0°C(±4.0°F)
Displayed room temperature range: 5.0°C ~ 59.9°C (41.0°F~139.8°F)
Room temperature accuracy:
±1°C (±2°F)
Measure time:
1 second
Measure interval:
1 second
Dimension:
130.1(L)x45.0(W)x50.3(D)mm
Weight:
About 62.0g (without battery)
Forehead and object operating
condition:
15.0°C~40.0°C (59.0°F~104.0°F)
with a relative humidity of 15%~85%
Atmosphere pressure: 700hPa-1060hPa
Storage and transport condition:
-20°C~55°C (-4°F~131°F) with a relative
humidity of 15%~90%
Atmosphere pressure: 700hPa-1060hPa
RF Frequency
2402 MHz - 2480 MHz
Modulation type
GFSK
Number of channel
40
Chanel spacing
2 MHz
Bluetooth version
V4.0
Maximum Output Power
8,31 dBm
Item
Group A1
Group A2
Group B
Group C
Clinical bias
∆
cb
-0,185
-0,175
-0,173
-0,182
Uncertainty
L
A
0,357
0,349
0,408
0,391
Clinical repeatability
σ
r
0,102
0,081
0,118
0,212
Summary of Contents for TH1009N-C