9.2 Changing the arm cuff
If you need a new arm cuff for your "Pressio A", this can be obtained from
Terraillon as an accessory.
For technical reasons, the arm cuff should only
be changed by Terraillon.
Please consult your dealer if necessary.
10 Descriptions of errors
10.1 Error numbers
Numbers
Causes
Err 0
No pulse or detected pulse is out of range.
Err 1
Arm cuff serious leakage or inflation too low.
Err 2
Systolic cannot be measured.
Err 3
Arm cuff pressure is too high.
Err 4
Diastolic cannot be measured or is outside of range.
Err 5
Deflation is too fast.
Err 6
Deflation is too slow.
Err 7
Systolic outsides the legal range.
Err 8
Mean pressure is outside of range.
Err 9
Arm cuff pressure is not enough even if the highest value is pumped.
10.2 Errors classifications and corrections
No display after push the
“START/STOP” button
Low battery symbol appears
The position of the batteries is
incorrect
The voltage of the batteries is
extremely low
The arm cuff is not fastened
Fasten the arm cuff again
The voltage of the batteries is
extremely low
When chilly, the batteries might
have a weak electron release
May be using wrong battery type.
Battery life is too short.
Use only Alkaline batteries.
The voltage of the batteries is
extremely low
No reading after measurement
Replace both batteries with
new ones
May not be in correct position
during measuring
The blood pressure naturally
varies from time to time
Suspicious measurement results.
Higher or lower than user’s
average measurement
Pulse rate higher/lower
than user’s average.
Always take measurements
in correct position (e.g 6)
Relax and try to
measure again
Do not move arm or body
during measurement
Do not measure right
after exercice
Push the
“START/STOP”button
again, and then
measure again
System design
This is normal: if user's blood
pressure exceeds the initial
pressure value, the device will
automatically pump to a higher
pressure, by 20mmHg each time,
until it reaches a suitable level.
Relax and try to
measure again
Power switches off automatically
Device reinflates during
measurement.
Move arm or body during
measurement
Measuring right after exercice
Insert the batteries in the
correct position (e.g 5.1)
Replace both batteries with
new ones.
Replace both batteries with
new ones.
Warm up the batteries,
or place in a warm place for
a while then re-start again
Conditions
Causes
Corrective Actions