Instructions for Use
Cleaning of the glass
The glass should be clean and free from stains in order
to ensure the quality of the indicator’s analysis.
Procedure:
1. Turn off the Trazanto Reader, and unplug the power
and USB cables from the device.
2. Wait until the glass is at room temperature.
3. Slightly moisten the included microfiber cloth with
isopropyl alcohol (95%). Verify that the cloth is clean.
Do NOT use water.
4. Open the cover and carefully clean the glass. Use gloves.
5. Wait a few minutes while the alcohol evaporates.
Verify that there are no stains or remains left on the glass.
6. Close the cover.
Do not clean the glass if the device is at a
temperature higher than room temperature,
since this could cause serious damages to
the device.