Sound Level: 

Approximately 48 dBA on low speed measure at 30 in. from the filter face, with the fan 

delivering an average air flow velocity of 90 FPM (.45 m/s). Ambient sound level less than 35 

dBA. 

 

Vibration Level: 

Will not exceed .09 MILS RMS. 

 

Housing: 

Both the fan plenum and filter housing have a mill finish (powder coat paint or stainless steel 

exterior optional), and sealed airtight. 

 

Prefilter: 

Furnished with a 16.0 in. x 23.25 in. x 0.25 in. polyurethane foam, washable prefilter. 

 

HEPA Filter: 

Factory tested and rated 99.99% efficient in removal of particles 0.3 micron and larger; leak 

free in accordance with the latest I.E.S.T. Recommended Practices. 

 

Filter Media: 

Microglass fiber with hot melt separators, sealed to the aluminum housing. 

 

Filter Face Guard: 

Epoxy powder coated expanded steel. 

 

Fan/Motor: 

Direct drive; forward curve centrifugal type with permanently lubricated sealed ball bearings. 

 

Motor: 

Permanent split capacitor type, rated for continuous duty furnished with thermal overload 

protection and a three-speed switch. 

 

Full-Load Amp: 

3.0 (for 120V, 60Hz, 1 phase); 1.5 (for 220V, 50/60Hz, 1 phase) Wattage Rating @ 90 fpm 

(0.45 m/s) – see chart below for other wattage ratings 

 

 

2 x 4 Models: 

160 Watts @ 90 fpm, 60Hz (145 @ 50Hz) 

 

2 x 3 Models: 

140 Watts @ 90 fpm, 60Hz (113 @ 50Hz) 

 

2 x 2 Models: 

72 Watts @ 90 fpm, 60Hz (67 @ 50Hz) 

 

To calculate sensible heat (in BTUH), multiply watts by 3.412 (e.g., for a 2' x 4' unit, at low speed, 170W x 3.412 = 

580 BTUH) 

 

System Options:

 

 

Lighting: 

Fluorescent type, approximately 1,000-foot candles; does not disturb the air flow. 

 

Ionizing Power  

110VAC, 50/60 Hz, single-phase; current draw: 0.3 amps (Cat. No. 1600-09A) 

 

Supply Ionizing Bars 

 

Cat. No. 1525-90  48"L brushed stainless steel bar, 8 emitters 

 

Cat. No. 1525-92  72"L brushed stainless steel bar, 12 emitters 

 

HEPA Filters 

 

The filtration medium consists of highly efficient microfiberglass fiber folded over corrugated separators of aluminum 

which prevent the media from nesting. This design channels air flow with optimal efficiency to reduce resistance. 

 

All filters are tested for air flow resistance and DOP (dioctylphthalate) smoke penetration (with a mean particle 

diameter of 0.3 microns) by measuring upstream and downstream smoke concentrations with a photometer. 

Summary of Contents for CleanBooth 1870-03B

Page 1: ...1800 73 CleanBooth Laminar Flow Station Copyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Revised May 2020 Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA...

Page 2: ...ment No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 2 Contents Contents 2 1 Introduction 3 2 Description 3 3 Installatio...

Page 3: ...microns with 99 999 efficiency Flexible anti static vinyl curtains are available to enclose the work area These curtains mount to the FFU support stand by means of a flexible GripTrack system System...

Page 4: ...1 TEL 714 578 6000 FAX 714 578 6020 4 3 Installation The CleanBooth is normally shipped with FFUs pre mounted in the support stand FFUs rest on the support frame no fasteners are required Optional cur...

Page 5: ...ight 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 60...

Page 6: ...n Copyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714...

Page 7: ...rra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 7 Step 4...

Page 8: ...Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 8 3 2...

Page 9: ...opyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 57...

Page 10: ...Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 10 Step 3 OPTIONAL Appl...

Page 11: ...rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 11 Step 5 Place the curtain pa...

Page 12: ...Flow Station Copyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578...

Page 13: ...ed monitor Under typical conditions 8 hours of operation per day filters should be replaced every year As a HEPA filter clogs you will notice increased blower noise and vibration along with decreased...

Page 14: ...g Frame and Curtains Periodically wipe down frame and optional curtain with a clean non linting wiper polyester recommended using clean water and a mild detergent if necessary Use unidirectional not c...

Page 15: ...ricated sealed ball bearings Motor Permanent split capacitor type rated for continuous duty furnished with thermal overload protection and a three speed switch Full Load Amp 3 0 for 120V 60Hz 1 phase...

Page 16: ...or separately For Terra s Policy and Procedures for Returning Goods see Terra s Internet site www TerraUniversal com Generally customers can improve the chance of collecting on a freight claim by foll...

Reviews: