Quick-Start Operating Guide 

 

 

High- CleanBooth™ Laminar Flow Station 

© 

Copyright 2020 Terra Universal Inc.  All rights reserved • Revised May 2020 • Document No. 1800-73

v2.0tm

 

Terra Universal, Inc.  •  TerraUniversal.com  •  800 S. Raymond Ave.  •  Fullerton, CA  92831  •  TEL: (714) 578-6000  •  FAX: (714) 578-6020 

14 

 

 

 

 

 

Fluorescent Tube   

 

Cat. # 

 

 

For 47” fixture 

 

 

3800-60 

 

 

For 35.5” fixture   

 

3800-61 

 

 

Replacement Fluorescent Fixture

   (including bulb) 

 

 

47” 

 

 

 

1870-11 

 

35.5” 

 

 

Ionizing Bar

 

 

The ionizing bar requires periodic replacement of its Tungsten emitter (recommended every two years). 

 

 

To replace emitter screw

, use 1/4-inch hollow socket to remove stainless steel emitter screw from the ionizing bar. 

Use care not to damage the whisker emitter when placing the socket over the screw head. Replace with unit below. 

 

Cat. # 

Tungsten Emitter 

1525-93 

 

Cleaning Frame and Curtains

 

 

Periodically, wipe down frame and optional curtain with a clean, non-linting wiper (polyester recommended), using 

clean water and  a mild detergent, if necessary. Use unidirectional (not circular) motion to remove surface 

contaminants from the edges of all surfaces. 

 

5.0

 

Specifications 

 

Overall Dimensions (see drawings for details) 

 

No. 1870-54 

53"W x 30"D (base)/28"D (top) x 63.5 / 91"H 

 

No. 1870-56 

77"W x 30"D (base)/28"D (top) x 63.5 / 91"H 

 

Construction 

 

Frame:  1" x 1" and 1" x 3" square CR steel tubing, powder-coated white 

 

FFU: 

Aluminum housing (see other FFU specs below) 

 

Performance:  Meets or exceeds the requirements/specifications of the following industry standards: 

 

ISO Class 4 (Class "100" (M 3.5)) air per the latest I.E.S.T. 14664-1 (Airborne Particulate Cleanliness Classes). 

 

I.E.S.T. Recommended Practices (Laminar Flow Clean Air Devices) Specifications 

 

Fan Filter Unit Specifications: 

 

Airflow Velocity: 

90 (±10) FPM (.45 m/s), when measured using a Velgrid (eight readings) 6 in. (152mm) 

below the filter face. 

 

 

Summary of Contents for CleanBooth 1870-03B

Page 1: ...1800 73 CleanBooth Laminar Flow Station Copyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Revised May 2020 Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA...

Page 2: ...ment No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 2 Contents Contents 2 1 Introduction 3 2 Description 3 3 Installatio...

Page 3: ...microns with 99 999 efficiency Flexible anti static vinyl curtains are available to enclose the work area These curtains mount to the FFU support stand by means of a flexible GripTrack system System...

Page 4: ...1 TEL 714 578 6000 FAX 714 578 6020 4 3 Installation The CleanBooth is normally shipped with FFUs pre mounted in the support stand FFUs rest on the support frame no fasteners are required Optional cur...

Page 5: ...ight 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 60...

Page 6: ...n Copyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714...

Page 7: ...rra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 7 Step 4...

Page 8: ...Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 8 3 2...

Page 9: ...opyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 57...

Page 10: ...Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 10 Step 3 OPTIONAL Appl...

Page 11: ...rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 11 Step 5 Place the curtain pa...

Page 12: ...Flow Station Copyright 2020 Terra Universal Inc All rights reserved Revised May 2020 Document No 1800 73v2 0tm Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578...

Page 13: ...ed monitor Under typical conditions 8 hours of operation per day filters should be replaced every year As a HEPA filter clogs you will notice increased blower noise and vibration along with decreased...

Page 14: ...g Frame and Curtains Periodically wipe down frame and optional curtain with a clean non linting wiper polyester recommended using clean water and a mild detergent if necessary Use unidirectional not c...

Page 15: ...ricated sealed ball bearings Motor Permanent split capacitor type rated for continuous duty furnished with thermal overload protection and a three speed switch Full Load Amp 3 0 for 120V 60Hz 1 phase...

Page 16: ...or separately For Terra s Policy and Procedures for Returning Goods see Terra s Internet site www TerraUniversal com Generally customers can improve the chance of collecting on a freight claim by foll...

Reviews: