PROBLEM:
The humidity module delivers an obviously incorrect humidity reading.
POSSIBLE SOLUTIONS:
1.
Check the low-voltage connection on the rear of the humidity module. If the connection to the sensor is
good, and the unit still fails to deliver an accurate reading, contact Terra Universal.
2.
The fan inside the humidity controller is a key part of the system and must be running to ensure
accurate RH readings. The system will not provide correct readouts if the fan is not running.
7.0
Specifications
Refer to the original order form for the exact specifications/configuration of your desiccator.
NOTE
Automatic RB (Relief/Bleed) Valve
Operating Temperature:
Dimensions:
0.5"W x 0.5"D x 1.5"H
Material:
Styrene-acrylonitrile (SAN) resin
Smart Relative Humidity (RH) Module
Operating Temperature:
Dimensions:
1.75”W x 5.5”D x 1.5”H
Case Material:
304 Stainless steel
Power Requirements:
120/240 VAC, 50/60Hz
Gas Inlet:
1/4” OD polyethylene tubing