Quick-Start Operating Guide

Series 100 Polymer Glove Box

© Copyright 2004 Terra Universal Inc. All rights reserved.

•    Revised December 2004    •    Document No. 1800-42

®

®

5

Terra Universal, Inc. • TerraUniversal.com • 700 N. Harbor Blvd. •  Anaheim, CA 92805-2553 • TEL: (714) 526-0100 • FAX: (714) 992-2179

Glove Box with Dual Purge System Only

1.

Make sure that your glove box is placed on a level, stable surface, away from heat or chemicals that could damage it.

2.

Using 1/4" tubing (not included), connect your in-house gas supply to the “Gas In” bulkhead on the rear panel of the Dual
Purge System by loosening the Poly-Tite collar, inserting the tubing, and tightening the collar.

3.

Use another length of 1/4" tubing to connect “Internal Pressure,” on the rear panel on the Dual Purge, to the enclosure to
be monitored.

4.

Connect the “Gas Out” port of the Dual Purge System to the inlet of your glove box.

5.

Connect the “Purge Control” ports of Dual Purge and NitroWatch systems with a 4-conductor sensor wire (terminated with
easy-snap telephone jacks). If your application has no NitroWatch, leave this port open.

6.

Plug the power cord to a grounded receptacle. The Dual Purge operates on 110VAC, 60Hz or 220VAC, 50/60Hz. No power
switch settings are required.

7.

Zero Calibration:

Press and hold SET and UP buttons while power is turned ON.

This activates a pressure sensor zero calibration routine before the setup mode is started. The display will indicate “CAL”
before this calibration is performed. The zero value is set automatically and is not changeable via the pushbuttons.

Remember to disconnect all air pressure from both pressure transducers before performing a zero calibration. Any
pressure on the transducers while this routine is run will be subtracted from the readings during normal operation,
resulting in likely continuous low-pressure alarms.

8.

Initial Programming: Turn ON the Dual Purge System while depressing SET (the left button on the front control panel).
Each time you release and then press and hold the SET button, you will advance through the following control functions. 
Use UP/DOWN to change default settings (shown in parentheses):

Prg1

Purge Delay (minutes): No. of minutes that the high purge remains active after the internal pressure level is
restored (default: 0).

Prg2

Purge Delay (seconds): No. of seconds that the high purge remains active after the internal pressure level is
restored (default: 60).

door

No. of seconds a door must remain open to activate OPEN DOOR alarm (defaults: 60). 

Inactive in Glove

Box applications

.

beeP

State of the BEEPER enable (default: ON).

Glo

Activates GLOVE BOX control mode – press UP to turn ON and proceed to glove box pressure setting; DOWN
to turn OFF (default: ON).

Press

In GLOVE BOX mode, this setting lets you adjust the pressure (milli-inches WG) at which the high purge is
activated (default: 0.2). 

Note:

complete the “Zero Calibration” procedure above (Step 6) before completing

this adjustment.

DonE

Settings are complete; press SET once more to begin operation.

Make a mistake? No Problem!

To reset the system, turn the system OFF while in setup mode and restart while

holding the SET button.

9.

Once programming is complete, turn the pressure regulator (the round knob on the right side of the Dual Purge control
panel) until the pressure gauge reads between 30 and 40 psi (use more pressure for larger glove boxes with 4 arm ports).
WARNING!!!: For safe operation, the incoming gas line must be regulated below 70psi.

10.

The Dual Purge System includes a flowmeter, which provides a continuous low-level purge to maintain a constant
positive pressure inside the cabinet. The flowmeter also ensures an uninterrupted nitrogen flow if power or system
electronics ever fail.

If continuous purging is desired, set the flowmeter so that the pressure gauge reads  at least 0.1" WG when the system is
in low purge state. This setting is recommended if you require a very low humidity set point (less than 10%RH) or if the
air lock doors will be opened frequently. Continuous purging will tend to drive the humidity level very low.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for 100 Series

Page 1: ...underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos In stock Ready to ship TAR certified secure asset solutions Expert team I Trust guarantee I 100 satisfaction All trademarks brand names and brands appearing herein are the property of their respective o...

Page 2: ... 42 Series 100 Polymer Glove Box Terra Universal Inc TerraUniversal com 800 S Raymond Ave Fullerton CA 92831 TEL 714 578 6000 FAX 714 578 6020 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 3: ...eoprene gaskets on the side doors and back panel provide a safe positive seal while allowing full access to the internal chamber to insert or remove materials or equipment The side doors open inward to allow an operator to pass materials in and out of the glove box easily The Series 100 Glove Box maintains a positive pressure for a clean low humidity environment It comes in five different material...

Page 4: ... response capacitive probe whose capacitance is proportional to the humidity level The sensor is capable of measuring humidity over the entire humidity range from 0 to 100 RH with an accuracy of 3 RH at a temperature of 68 degrees F The voltage output of the sensor is directly proportional to the humidity A 3 prong connector allows easy connection of the sensor to the Control Unit The Control Unit...

Page 5: ... chamber either 100 no recirculation 0 complete recirculation or 50 half fresh half recirculated See Catalog No 105 for additional information Quick Start Operating Guide Series 100 Polymer Glove Box Copyright 2004 Terra Universal Inc All rights reserved Revised December 2004 Document No 1800 42 3 Terra Universal Inc TerraUniversal com 700 N Harbor Blvd Anaheim CA 92805 2553 TEL 714 526 0100 FAX 7...

Page 6: ...in use Air Locks Air Locks can be mounted on either side of the chamber to provide a safe practical way of moving materials into and out of the work area Materials are loaded into the air lock by means of an external door an operator then opens an internal door inside the glove box to bring materials into the work area Each air lock features its own gas inlet system pressure and an Automatic RB Va...

Page 7: ...ease and then press and hold the SET button you will advance through the following control functions Use UP DOWN to change default settings shown in parentheses Prg1 Purge Delay minutes No of minutes that the high purge remains active after the internal pressure level is restored default 0 Prg2 Purge Delay seconds No of seconds that the high purge remains active after the internal pressure level i...

Page 8: ...tial NitroWatch Programming After connecting the NitroWatch to the Dual Purge System turn ON the Dual Purge System while holding down the NitroWatch SET button the left button on the front control panel Each time you release and then press and hold the SET button you will advance through the following control functions Use the UP DOWN buttons to change default settings shown in parentheses Door Op...

Page 9: ... Water Gauge WG The Dual Purge System includes a flowmeter which provides a continuous low level purge to maintain a constant positive pressure inside the cabinet The flowmeter also ensures an uninterrupted nitrogen flow if power or system electronics ever fail If continuous purging is desired set the flowmeter so that the pressure gauge reads at least 0 1 WG when the system is in low purge state ...

Page 10: ...warm water with or without a mild non abrasive detergent and a dry clean non abrasive wipe Use only light pressure when cleaning If the outside surface is exceptionally dirty or gritty lightly swab the surface with a saturated cloth and allow the surfactants to drain away Avoid rubbing dirt or grit into the surface Turn the cloth often and replace with a clean cloth frequently Dry the surface by b...

Page 11: ...ave programmed the Dual Purge for Glove Box service and that the internal pressure setting is between 0 2 and 0 7 inches water gauge 2 If your system incorporates a NitroWatch and you require a very low humidity level you may need to increase the flowmeter setting on the Dual Purge System to provide more than 0 1 WC internal pressure You should increase the flowmeter setting until the system maint...

Page 12: ...1 4 clear plexiglass or optional material you ve selected with white translucent bottom 8 x 8 door openings with gasketed doors on both sides of unit 8 round arm ports with wide flanges for connecting glove assembly Tapped 1 8 NPT inlet for connecting the Glove Box Dual Purge System and 1 8 NPT system pressure port Note Stainless steel work surface is recommended on all polymer glove boxes to prev...

Page 13: ... from misuse intentional or unintentional abuse neglect natural disasters or acts of God Products Manufactured by Others Terra Universal Inc warrants that to the best of its ability Terra s representations of products which are manufactured by others reflect the manufacturer s representations subject to change without notice Sole warranty for these products is the original manufacturer s warranty ...

Page 14: ...d sleeve assembly into the glove box allowing 2 of sleeve to extend beyond the front edge of the mounting flange 2 Fold the outside edge of the sleeve over the flange so that roughly 2 of the sleeve cuff is stretched along the outside perimeter of the flange 3 The sleeve is held in place by a stainless steel clamp TUI Cat No 1689 40 Position this clamp along the mounting flange so that about 1 of ...

Page 15: ...quipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentation Artisan Scientific Corporation dba Artisan Technology Group is not an affiliate representative or authorized distributor for any manufacturer listed herein We re here to make your life easier How...

Reviews: