3.1    Measure Voltage/Display Mode 

 

Now it will be ask for your choose off the measure voltage. 
In the display appears : 

VOLTGE MODE : 
AUTO   

(MAN.10V)    (MAN. 100V) 

 

By pressing key (B) 

Change

 the display is switching between AUTO (Automatic), MAN. 10V, and MAN. 

100V. Press the (A) 

Set-Up

 key to choose the displayed voltage mode. 

 
After that, it will be ask for your choose off the display voltage. 
 

DISPLAY MODE : 
R>2.0xE12?   

(R>2.0 T? ) 

 

By pressing key (B) 

Change

 the display is switching between the mathematical and technical displa y form. 

Press the (A) 

Set-Up

 key to choose the displayed display time. 

 
Now the unit will execute a reset, and in the display appears if timer is on, else go to point 3.2 Timer off. 
 

T=xxxs Avg Auto   

(Lap)    (+10V, +100V) 

PRESS START ! 
 

By pressing key (A) 

Start

 the next measurement will be started. 

The timer decrements in 1 second steps up to Zero, then the measure value will appears in the display as 
R

avg 

by AVERAGE - Mode and R

lap 

by Lapse 

 Mode. 

In the 2nd line the air temperature and the humidity is displayed: 
 

Ravg = 2.5xE9?   

(Rlab = 2.5 G? ) 

AIR: xx

°

C xx%r.F 

 

Alternates all 2 second in the 2nd line appears: 
 

FN:xxx T=xxxs 

 

By pressing key (A) 

Start

 the measure value together with the air temperature and the Humidity will be 

saved in the EEPROM using the displayed File 

 No. These values can be read later with the PC - software. 

 
 

3.2    Timer 

 

If the timer is not activated, the TOM 600 directly switches on the resistance display in the on-line operation. 
According to the chosen display mode and voltage mode in the display appears: 
Example:   

R=5.5xE10W T=OFF   

   

AUTO 

AIR: xx

°

C xx%r.F

 

 
 
 
 
 
 

Summary of Contents for TOM 600

Page 1: ...surement High resistances up to 2x1012 Ohm 2 TERA Ohm can be measured with a voltage of 100 volts and a precision of 5 As peculiarity the ambient temperature and the humidity will be measured and stored to every measured resistance value TOM600 ME KLEINWACHTER Forschungs Entwlcklungs Produktlons u Vertrlebsges m b H 1 ...

Page 2: ...ed by switching off the unit The values can be read out trough the PC Interface at all times Since the resistance values depend very strongly on the Humidity and the air temperature these influences are measured together with the resistance value So a reproducible measure ment is given 2 1 Start Operation The TERA Ohmmeter TOM 600 is delivered with 4 x Mignon AA NiMH Batteries and is directly read...

Page 3: ...wer supply to a power socket If the appliance is switched off and the charging work correct the green Charge LED shines The NiMH batteries are completly charged after maximum of 14 hours and good for 10 hours continuous operation A longer charging time can dammage the batteries 2 3 Measure Range and Measure Voltage The TOM 600 has an automatic ranging approve the whole resistor range The appliance...

Page 4: ...ond steps up to 240 seconds Press the A Set Up key to choose the displayed timer time At the display appears TIMER MODE AVERAGE LAPSE By pressing key B Change the timer is switching between AVERAGE and LAPSE Press the A Set Up key to choose the displayed timer mode Now it is asked whether you will delete the EEPROM stored data files In the display appears DEL FILES YES NO By pressing key B Change ...

Page 5: ...off T xxxs Avg Auto Lap 10V 100V PRESS START By pressing key A Start the next measurement will be started The timer decrements in 1 second steps up to Zero then the measure value will appears in the display as Ravg by AVERAGE Mode and Rlap by Lapse Mode In the 2nd line the air temperature and the humidity is displayed Ravg 2 5xE9 Rlab 2 5 G AIR xx C xx r F Alternates all 2 second in the 2nd line a...

Page 6: ...t the current measurement can be completed If the Battery Voltage is lower as 4 3 volts the following display appear and the unit automatically switches off to save the batteries LOWEST BATTERY AUTO SWITCH OFF 4 1 Display Contrast On the front side left from the PC Interface there is a hole trough this you can screw a potentiometer for adjust the Contrast of the LCD Display System requirement l PC...

Page 7: ...nit is not approved for the use in power utilities Info Make sure that there is no voltage on your measuring External Voltage gives you a wrong measure value and can destroy your instrument Housing Plastic Housing 223 x 106 x 60 mm L x B x H Weight app 350 g Display alphanumeric Display 2 lines with 16 Letters Display field 60 mm x 25 mm Measure Ranges Resistor 20 kΩ 2 0 TΩ Tolerance 5 Temperature...

Page 8: ...ding Resistors and point topoint resistance specially for installed ESD Table or floor coverings l Ring Electrode to EOS ESD S 11 11 DIN EN 61340 5 1 2 3 comprising to the measurement of transit and surface resistances 1 shielded and 1 no shielded Measurecord The appliance is maintenance free If the appliance should get dirty it can be cleaned with a lint free cotton cloth with solvent free deterg...

Page 9: ...erface 3 Measure Voltage Output 4 Measure Input 5 Contrast Potentiometer for LC Display 6 Temperature Humidty Sensor 7 Charge control LED 9 12 Legend B ON OFF Charge OUT IN Rx Charge Battery Enter A START Set Up TERA Ohmmeter TOM 600 3 4 1 7 6 2 5 ...

Page 10: ...r die elektromagnetische Verträglichkeit Richtlinie des Rates vom 3 5 1989 89 336 Stand Juli 1993 erfüllt insbesondere der Normen EN 61010 Safety EN 55011 Gruppe 1 Klasse B EN 50082 2 92 für ISM Geräte Die Firma Kleinwächter GmbH hält folgende technischen Dokumentationen zur Einsicht bereit vorschriftsmäßige Bedienungsanleitung Baupläne technische Dokumentationen Hausen den 09 05 2004 Jürgen Brunn...

Reviews: