6. The lowest current for measuring is 10% of the nominal value. It is
important to consider that the measurement errors increase while test
current decreases. The equipment accuracy is specified for test
currents higher than 80%.
7. The display will show the resistance value measured and the
corresponding unit ( [ohms], m [milli-ohms] or µ [micro-ohms]).
8. The value can be retained in the display by pressing the hold key.
Pressing this key again, the value will be released.
9. Press the Stop key in order to finish the measurement. Do not turn
Off the equipment without pressing the Stop key before.
10. Finally, when finishing measurements, turn the equipment Off using
the On/Off switch.
Precaution: Do not connect or disconnect the test leads during the
measurement.
5. Messages
TENTECH
MPK-254
When turning the equipment On using the On/Off switch, this
introductory message appears for a while.
WAIT...
This message appears each time the equipment needs to adjust any
parameter in order to optimize the readings.
PRESS START
The equipment is ready to start a measurement, thus the operator has to
press the Start key.