TENSITRON LX-100-1 Operating Instructions Manual Download Page 8

       LX-1 Series Instrument           

www.tensitron.com                      

pg. 8

 

3.  Log Data: 

a.  Press the STORE button to store and display up to five 

separate tension readings along with the average of these 
values. 

b.  Clear the data displayed on the Main Screen by pressing 

and holding the STORE button, or power the instrument off. 

c. 

For additional information on logging and viewing data, refer 
to Sec 6 Display Screens - Data Logging. 

4.  Read Tension:  Variations in materials and material diameters affect 

tension readings. It is essential to select the correct material and size 
before use, or tension values may be incorrect. 

5.  Select Tension Units: 

a.  Using the up or down arrows, scroll to TENSION UNITS, 

then press ENTER. 

b.  Next, select from: Kilograms, DecaNewtons, or Pounds and 

then press ENTER. 

6.  Select Material:  

a.  Using either the up or down arrows, scroll to SELECT 

MATERIAL, then press ENTER. 

b.  Next scroll through the material descriptions until the correct 

material size is highlighted, then press ENTER. Your main 
display will indicate the material size selected. (See Figure 
4) 

c. 

If your specific material is not listed, add it to the menu by 
following the calibration instructions included at the end of 
these instructions. Or, send a 1

5’ sample of your material 

and the instrument to Tensitron and ask to have it added to 
the menu. 
 

                                

 

7.  Zero Instrument: Hold the instrument in the attitude your reading will 

be taken before you engage it to the tensioned material. Then, press 
ZERO. 

 

Engaging Instrument onto Cable 

W

ARNING

D

O NOT OVERTORQUE THE ENGAGEMENT LEVER OR YOU WILL DAMAGE THE INSTRUMENT

D

O NOT EXCEED THE MAXIMUM TENSION RANGE OF THE INSTRUMENT OR DAMAGE WILL OCCUR

Select Material Size 

– 

Instrument automatically 
adjusts calibration 

Figure 4:  

Select Material Screen 

Summary of Contents for LX-100-1

Page 1: ...LX 1 Series Instrument www tensitron com pg 1 Operating Instructions LX 1 Series Digital Wire Cable and Textile Tension Meter...

Page 2: ...uments 6 5 BASIC OPERATION 7 Charging Instrument Batteries 7 Quick Start Instructions 7 Engaging Instrument onto Cable 8 6 DISPLAY SCREENS 10 SETUP Screen 10 o LCD Refresh Rate 10 o Data Averaging 10...

Page 3: ...assistance please call us at 303 702 1980 or find us online at www tensitron com 1 WARRANTY POLICY STANDARD EQUIPMENT WARRANTY Tensitron warrants that all Tensitron manufactured equipment will be fre...

Page 4: ...ry design Battery Life Approximately 12 hours depending upon backlighting intensity Operating Temperature 32 F to 120 F If these values are exceeded battery charge discharge rates will decline However...

Page 5: ...maged instrument Tag damaged instrument DO NOT USE until repaired For repair send instruments to Tensitron o Maintain instrument carefully Protect instrument from exposure to water Keep instrument dry...

Page 6: ...ed readings o Data Statistics Monitor your process by viewing count average minimum maximum and standard deviation values of your logged data o Selectable Tension Units Choose to display tension in Lb...

Page 7: ...k Start Instructions 1 Power unit on by pressing ON button Main display will indicate Tension Material five stored tension readings and their average and other information See Figure 3 2 Select a scre...

Page 8: ...Material a Using either the up or down arrows scroll to SELECT MATERIAL then press ENTER b Next scroll through the material descriptions until the correct material size is highlighted then press ENTE...

Page 9: ...f it takes a lot of force to close the lever STOP Something is binding or the material is too large in diameter Closing the lever in this situation will damage the instrument Note the tension reading...

Page 10: ...readings before posting the average of those 3000 readings to the display Options are 1 2 5 and 10 seconds This is a rolling average meaning that only the readings taken in the last 1 2 5 or 10 second...

Page 11: ...on or off or to edit your high and low values See Figure 7 o Version Press this button to determine what model the instrument is set at in the firmware as well as to see what level of firmware is down...

Page 12: ...rds on correctly sized materials See Figure 8 4 Never use any tensioning device that attempts to convert rotational torque values into tension loads as these types of systems are highly inaccurate Als...

Page 13: ...ss ENTER The screen will now display SETUP RE CALIBRATE TENSION ENTER PASSWORD CAL DATA WILL BE LOST 2 Password To proceed with calibration you must enter the password You may obtain the password by e...

Page 14: ...BER OF POINTS __ ZERO UNIT Without any material engaged to the instrument hold it in the attitude you will be taking the reading in and press ENTER to calibrate zero weight Next the display will promp...

Page 15: ...LX 1 Series Instrument www tensitron com pg 15...

Page 16: ...733 S Bowen Street Longmont CO 80501 USA Phone 303 702 1980 Fax 303 702 1982 E mail service tensitron com Web Site www tensitron com...

Reviews: