TENSITRON LX-100-1 Operating Instructions Manual Download Page 12

       LX-1 Series Instrument           

www.tensitron.com                      

pg. 12

 

 

Scroll to VIEW DATA LOG and press ENTER. 

o

 

Clear Single Point Log 

To clear stored data, scroll to CLEAR SINGLE PT LOG, press ENTER, 
and follow the prompts. 

 

 

TENSION UNITS Screen 

To change the unit of measure in which tension readings are logged: 

 

Scroll to Tension Units and press ENTER. 

 

Use the Up and Down arrows to select Kilograms, 
DecaNewtons or Pounds and press ENTER. 

 

SELECT MATERIAL Screen 

Press ENTER while this display is shown to show the available options for the 
material types. Use the up or down arrow to scroll to the option you want and 
press ENTER to select it. 

7.    ADVANCED OPERATION 

 

General Calibration Precautions 

1.  To verify the accuracy of your instrument, check the tension on a sample 

length of material by suspending known weights from the opposite end. 

2.  When performing this test, use a micrometer to measure your material 

diameter/thickness and ensure it is dimensionally within tolerance, +/- 
.003” of size, and not out of round. 

3.  Only use free-hanging weights which are traceable to National Standards 

on correctly-sized materials. (See Figure 8.) 

4.  Never use any tensioning device that attempts to convert rotational torque 

values into tension loads, as these types of systems are highly inaccurate. 
Also, any load cell system must be routinely checked for accuracy using 
traceable weights as these types of systems are also highly inaccurate 
and generally do not meet the minimum repeatable accuracy standards 
required by ISO-17025. 

5.  NEVER SIMULATE TENSION LOADS ON THESE INSTRUMENTS BY 

USING A CALIBRATION BAR!  

 

Check Accuracy 

1.  Confirm that the size material 

you’re working with is the same size 

selected in your LX

’s display. 

2.  Before engaging the LX instrument onto the material, zero the instrument 

by holding 

it in the same attitude you’ll be taking the readings and press 

the “Enter” or “Zero” button. This will eliminate any small tare effects. 

3.  Take readings along the material. 

FOR

 

BEST

 

RESULTS

 

ALWAYS

 

TAKE

 

SEVERAL

 

READINGS

 

OF

 

THE

 

CABLE

 

TENSION

 

BY

 

DISENGAGING

 

AND

 

THEN

 

RE-ENGAGING

 

THE

 

INSTRUMENT

 

ONTO

 

THE

 

CABLE.

 

4. 

Verify the tension value. For example, if your suspended weight is 40.0 lb. 
your instrument should also indicate 40.0 lb. +/- 5.0 lb. once properly 
engaged to the material. 

This procedure confirms the instrument’s 

accuracy and it is now ready for use. 

 

Summary of Contents for LX-100-1

Page 1: ...LX 1 Series Instrument www tensitron com pg 1 Operating Instructions LX 1 Series Digital Wire Cable and Textile Tension Meter...

Page 2: ...uments 6 5 BASIC OPERATION 7 Charging Instrument Batteries 7 Quick Start Instructions 7 Engaging Instrument onto Cable 8 6 DISPLAY SCREENS 10 SETUP Screen 10 o LCD Refresh Rate 10 o Data Averaging 10...

Page 3: ...assistance please call us at 303 702 1980 or find us online at www tensitron com 1 WARRANTY POLICY STANDARD EQUIPMENT WARRANTY Tensitron warrants that all Tensitron manufactured equipment will be fre...

Page 4: ...ry design Battery Life Approximately 12 hours depending upon backlighting intensity Operating Temperature 32 F to 120 F If these values are exceeded battery charge discharge rates will decline However...

Page 5: ...maged instrument Tag damaged instrument DO NOT USE until repaired For repair send instruments to Tensitron o Maintain instrument carefully Protect instrument from exposure to water Keep instrument dry...

Page 6: ...ed readings o Data Statistics Monitor your process by viewing count average minimum maximum and standard deviation values of your logged data o Selectable Tension Units Choose to display tension in Lb...

Page 7: ...k Start Instructions 1 Power unit on by pressing ON button Main display will indicate Tension Material five stored tension readings and their average and other information See Figure 3 2 Select a scre...

Page 8: ...Material a Using either the up or down arrows scroll to SELECT MATERIAL then press ENTER b Next scroll through the material descriptions until the correct material size is highlighted then press ENTE...

Page 9: ...f it takes a lot of force to close the lever STOP Something is binding or the material is too large in diameter Closing the lever in this situation will damage the instrument Note the tension reading...

Page 10: ...readings before posting the average of those 3000 readings to the display Options are 1 2 5 and 10 seconds This is a rolling average meaning that only the readings taken in the last 1 2 5 or 10 second...

Page 11: ...on or off or to edit your high and low values See Figure 7 o Version Press this button to determine what model the instrument is set at in the firmware as well as to see what level of firmware is down...

Page 12: ...rds on correctly sized materials See Figure 8 4 Never use any tensioning device that attempts to convert rotational torque values into tension loads as these types of systems are highly inaccurate Als...

Page 13: ...ss ENTER The screen will now display SETUP RE CALIBRATE TENSION ENTER PASSWORD CAL DATA WILL BE LOST 2 Password To proceed with calibration you must enter the password You may obtain the password by e...

Page 14: ...BER OF POINTS __ ZERO UNIT Without any material engaged to the instrument hold it in the attitude you will be taking the reading in and press ENTER to calibrate zero weight Next the display will promp...

Page 15: ...LX 1 Series Instrument www tensitron com pg 15...

Page 16: ...733 S Bowen Street Longmont CO 80501 USA Phone 303 702 1980 Fax 303 702 1982 E mail service tensitron com Web Site www tensitron com...

Reviews: