21. EMC PRECAUTIONS

Use  special  precautions  regarding  EMC  according  to  the 
information provided below.
Other  portable  and  mobile  RF  communications  equipment  can 
affect performance. Do not use when adjacent to or stacked with 
other  electrical  equipment.  Use  of  leads  or  probes  other  than 
those listed in section 18 may affect EMC performance.

      
   

Guidance and manufacturer’s declaration - electromagnetic 
emissions

The Touch TENS is intended for use in the electromagnetic 
environment specified below.
The customer or the user of the Touch TENS should assure that it is 
used in such an environment.

Emissions test 

Compliance 

Electromagnatic environment - guidance

RF emissions 

Group 1 

The model itouch uses RF energy only for its

CISPR 11 

 

internal  function. Therefore, its RF emissions

 

 

are very low and are not likely to cause any

 

 

interference in nearby electronic equipment

RF emissions 

Class B 

The model itouch is suitable for used in

CISPR 11 

 

domestic establishment and in establishment

 

 

directly connected to a low voltage power

Harmonic emissions 

Not applicable 

supply network which supplies buildings used

IEC61000-3-2 

 

for domestic purposes.

Voltage fluctuations/ flicker  Not applicable
emissions IEC 61000-3-3

21

22

The Touch TENS is intended for use in the electromagnetic 
environment specified below. The customer or the user of the 
Touch TENS should assure that it is used in such an environment.

Immunity 

IEC 60601 test  Compliance  Electromagnetic environment – guidance

test 

level 

level

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any

 

 

 

part of the model itouch, including cables,

 

 

 

than the recommended separation distance

 

 

 

calculated from the equation applicable to

 

 

 

the frequency of the transmitter

 

 

 

Recommended separation distance 3V

 

 

 

d = 1.2√ P 80M Hz to 800MHz

Conducted  3 Vrms 

3 Vrms 

d = 2.3√ pt P 800MHz to 2.5GHz

RF 

150 kHz to 80 

 

where P is the maximum output power

IEC 61000-  MHz 

 

rating of the transmitter in watts (W)

4-6 

 

 

according to the transmitter manufacturer

 

 

 

and d is the recommended separation

 

 

 

distance in meters (m).

Radiated RF  3 V/m 

3 V/m 

Field strengths from fixed RF transmitters, 

IEC 61000-4  80 MHz to 2,5 

 

as determined by an electromagnetic site

-3 

GHz 

 

survey, should be less than the compliance

 

 

 

level in each frequency range.

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the symbol on the

 

 

 

left of this box

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the model itouch is used 
exceeds the applicable RF compliance level above, the model itouch should be observed
to verify normal operation. If abnormal performance is observed, additional measures may 
be necessary, such as re-orienting or relocating the model itouch.
b.Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Summary of Contents for TouchTENS XL-T

Page 1: ...Model XL Y1 T INSTRUCTIONS FOR USE...

Page 2: ...lip 4 7 The Programmes 4 8 Programme Settings 6 9 Controls and Display 7 10 Controls 8 11 Connecting Lead wires 11 12 Positioning of Electrode Pads 12 13 General Pad Advice 14 14 Cleaning 14 15 Troubl...

Page 3: ...energy output levels for easy use A choice of eight pre set programmes is available The operation of the controls is described in detail in Sections 9 10 The Tenscare Touch TENS is a dual channel TENS...

Page 4: ...ispose of batteries safely Do not throw batteries onto a fire Risk of explosion Battery Life Batteries should last about 48 hours at full power Unused batteries have a nominal shelf life of 3 years bu...

Page 5: ...00 Modulated modulating continuously this gives a pleasant slow relaxing massage sensation G Frequency This changes the frequency every 3 minutes Pain Gate 2 110 200 Modulated stepping from 2 to 110 H...

Page 6: ...re are four coloured buttons STRENGTH CONTROLS ON OFF Channel 1 Strength DOWN Channel 1 Strength UP Channel 2 Strength UP Channel 2 Strength DOWN Time Selected Programme selector Channel 1 Strength Ch...

Page 7: ...that the unit will continue functioning until you switch it off You can set session times of 20 40 or 60 minutes by pressing the green button repeatedly The LCD shows the session duration next to the...

Page 8: ...ee Section 16 for information on where you should not place pads 3 Elapsed Time Press Programme and Right blue together and hold down for 3 seconds The display will show U for Usage and a number from...

Page 9: ...of the tackiness of the pad by adding a tiny drop of water on each pad and spreading around If too much water is added the pads will become too soft then it is suggested in order to try re establish...

Page 10: ...being connected to monitoring equipment with body worn electrode pads it may interfere with the signals being monitored Strong electromagnetic fields electrosurgery microwave cookers mobile phones may...

Page 11: ...This will invalidate the guarantee Mix old new or different types of batteries Use this device leads electrodes and accessories other than those recommended by the manufacturer Performance may vary fr...

Page 12: ...f the problem The guarantee does not cover accessories the batteries electrode pads or lead wires Please note that the Guarantee is invalidated if incorrect batteries have been fitted or the unit has...

Page 13: ...e model itouch including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance 3V d 1 2 P 80M Hz t...

Page 14: ...rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the max...

Page 15: ...TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Distributed by Pub No I XLY1T UK Version 1 1 06 13 Manufactured by...