36

35

EMC PRECAUTIONS

Use special precautions regarding EMC according to the information  
provided below.
Other portable and mobile RF communications equipment can affect performance. 
Do not use when adjacent to or stacked with other electrical equipment. Use of 
leads or probes other than those listed in section 18 may affect EMC performance.

The Itouch Plus is intended for use in the electromagnetic environment specified  
below. The customer or the user of the Itouch Plus should assure that it is used in  
such an environment.

Immunity test   

IEC 60601 test level   

Compliance level 

Electromagnetic  environment guidance

Electrostatic 

±6 kV contact 

±6 kV contact    

Floors should be wood, concrete or  

discharge (ESD) 

±8 kV air 

±8 kV air 

ceramic tile. If floors are covered with  

IEC 61000-4-2   

 

 

synthetic material, the relative

 

 

 

humidity should be at least 30 %.

Electrical fast  

±2 kV for power supply  Not applicable 

Not applicable 

transient/burst 

lines 

IEC 61000-4-4 

±1 kV for Input/output lines

Surge       

±1 kV differential mode  Not applicable 

Not applicable 

IEC 61000-4-5   

±2 kV common mode

Voltage dips, short 

<5 % UT (>95 % dip in   Not applicable  

Not applicable 

interruptions and 

UT) for 0,5 cycle 

voltage variations 

40 % UT(60 % dip in  

on power supply 

UT ) for 5 cycles 

input lines 

70 %UT  (30 % dip in   

IEC 61000-4-11 

UT )for 25 cycles  

 

<5 %  UT  (>95 % dip  

 

in  UT ) For 5s  

 

Power frequency 

3 A/m 

Not applicable 

Not applicable 

(50/60 Hz) magnetic  
field IEC 61000-4-8

NOTE   UT is the a.c. mains voltage prior to application of the test level

Guidance and manufacturer’s declaration - electromagnetic emissions

The itouch Sure is intended for use in the electromagnetic environment specified below.  
The customer or the user of the itouch sure should assure that it is used in such an environment.

Emissions test 

Compliance 

Electromagnatic environment - guidance

RF emissions 

Group 1

 

The model itouch uses RF energy only for its  

CISPR 11

 

 

internal  function. Therefore, its RF emissions are very  

 

 

low and are not likely to cause any interference in  

 

 

nearby electronic equipment.

RF emissions 

Class B

 

The model itouch is suitable for used in domestic

CISPR 11

 

 

establishment and in establishment directly connected  

 

 

to Harmonic emissions a low voltage power supply  

Harmonic emissions 

Not applicable

 

network which supplies buildings used for domestic  

IEC61000-3-2

 

 

purposes.

 

Voltage fluctuations/ 

Not applicable 

flicker emissions 

IEC 61000-3-3 

 

 

Summary of Contents for itouch easy

Page 1: ...c t i o n s f o r U s e I N T E L L I G E N T T E N S TensCare Ltd 9 Blenheim Road Longmead Business Park Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 Email painaway tenscare co uk www tenscare co uk...

Page 2: ...Screen Information 9 8 Contents 15 9 Installation of Batteries 15 10 Connecting Mono Lead wire 17 11 Positioning of Electrode Pads 18 12 General Pad Advice 20 13 Pad Alarm 22 14 The Belt Cradle 23 15...

Page 3: ...th CSC you can increase the strength on the itouch easy in very small steps which is much more comfortable than conventional TENS units One Touch Memory Start The itouch easy has a memory which allows...

Page 4: ...ificant reduction in pain if not complete pain relief 4 SIDE EFFECTS There are no known side effects to TENS use and long term TENS use is not harmful 5 THE PROGRAMMES There is no one programme approp...

Page 5: ...sation Specific Uses E Constant This is similar to A but has Most types of acute deeper penetration and a and chronic pain stronger feel It is more including arthritis likely to produce muscle sciatic...

Page 6: ...his enables the body to receive many different signals and can be very beneficial and lessen any effect of accommodation See programmes F and G Han Stimulation Professor Han in China found that altern...

Page 7: ...ton Channel 2 Strength UP Channel 2 Strength DOWN Channel 1 Strength UP Time Selector Channel 2 Strength UP 7 SCREEN INFORMATION Programme selected Frequency selected Channel 1 strength selected Pulse...

Page 8: ...to switch the unit off press and hold down this button for 3 seconds ON Channel 1 Strength UP button Note Channel 1 controls the output through the white leads To switch unit on Press and hold down fo...

Page 9: ...press the Programme Selector button and release The programmes are shown as A B C D E F G or H Time Selector button Each time the unit is switched on it defaults back to continuous mode known as TIMER...

Page 10: ...l Ensure that the batteries are inserted the right way as shown in battery compartment When the batteries are running low a low battery indicator will show on the screen battery symbol and it is impor...

Page 11: ...the other end of the Mono Lead wire the lead wire splits into four pin ends 2 white and 2 black Push the pin ends into the pigtail ends of the electrode pads The mono lead wire can be split as requir...

Page 12: ...rotective plastic shield by holding and lifting one corner of the pad and pulling Do not pull on the pigtail wire of the pad After use always replace the pads on the plastic liner and replace in the r...

Page 13: ...contact between any of the pads and the skin or the quality of the pads is below the required standard the pad alarm will be triggered An audible bleep will be heard and at the top of the screen the...

Page 14: ...ng solution than soap and water Clean the case and lead wires at least once a week by wiping with a damp cloth and a solution of mild soap and water Wipe dry 16 TROUBLESHOOTING Problem Possible causes...

Page 15: ...S may affect seizure threshold During the first three months of pregnancy It is not known whether TENS may affect foetal development On the abdomen in the later stages of pregnancy Stop use immediatel...

Page 16: ...brain are unknown Near malignant tumours In vitro experiments have shown that electricity can promote cell growth Do not Ignore any allergic reaction to the electrode pads If a skin irritation develo...

Page 17: ...h 19 CONSUMABLES AND SERVICING Replacement electrode pads new batteries and lead wires are available from your supplier or distributor see back cover for contact details by mail order from TensCare by...

Page 18: ...ked with a crossed out wheelie bin symbol as depicted below To comply with the Directive you can return your old electro therapy unit to us for disposal Simply print a postage paid PACKETPOST RETURNS...

Page 19: ...ould be protected from liquids NB The electrical specifications are nominal and subject to variation from the listed values due to normal production tolerances EC Declaration of Conformity TensCare Lt...

Page 20: ...5 UT 95 dip in Not applicable Not applicable interruptions and UT for 0 5 cycle voltage variations 40 UT 60 dip in on power supply UT for 5 cycles input lines 70 UT 30 dip in IEC 61000 4 11 UT for 25...

Page 21: ...marked with the following symbol Rated maximum output power of transmitter W NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations El...

Page 22: ...39 NOTES...

Reviews: