13
and intensity cannot be increased.
Check electrode pad position. If you are certain that the pads are positioned safely,
press the SKIP key
to override. The triangle will stop flashing and intensity can be
increased to 60mA.
Note that the two channels ADD intensity, so that
max effective IF intensity is 120mA.
Memory Mode
Pressing the MODE button enters Memory Mode
The Usage time will be accumulatively recorded when the output level is
above zero. The accumulative treatment time in minutes and the number of
uses are displayed.
Press Mode key again to return back to the previous normal display
Pressing “CH2▼” key and “Mode“ key together for 3 seconds will clear the treat
ment
time to zero.
When a programme is running, this key also acts a PAUSE button. The PAUSE symbol II
will be displayed, and the programme timer will stop. Pressing again resumes the
programme and the intensity gradually returns to the set value.
SKIP Key
High intensity I.F. can be harmful if electrodes are placed so that the current
goes directly through the chest or across the head. For your safety, when
intensity reaches 40mA, the warning triangle flashes, and intensity cannot
be increased.
Manual Programme Lock
When “Manual” is showing, you can protect the manual settings by pressing
and holding this key for 3 seconds.
The key symbol will flash, and you will be unable to change the manual
settings.