![Tens i-Pulz 2 User Manual Download Page 31](http://html1.mh-extra.com/html/tens/i-pulz-2/i-pulz-2_user-manual_1087591031.webp)
31
i-Pulz 2
Table 2. Immunity Guidance for Users
Guidance and Manufacturer’s declaration -
electromagnetic emissions
The i-Pulz 2 (T1) is intended for use in the electromagnetic environment
specified below. The customer or the user of i-Pulz 2 (T1) should assure that it
is used in the following environment:
Immunity Test
IEC 60601-1-2
Test Level
Compliance
level
Electromagnetic
Environment -
guidance
Electrostatic
discharge (ESD)
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient / burst
Not applicable
The device is battery
powered
Surge
Not applicable
The device is battery
powered
Voltage
dips, short
interruptions,
and voltage
variations on
power supply
input lines
Not applicable
The device is battery
powered
Power
frequency
(50/60 Hz)
magnetic field
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Summary of Contents for i-Pulz 2