![Tens i-Pulz 2 User Manual Download Page 26](http://html1.mh-extra.com/html/tens/i-pulz-2/i-pulz-2_user-manual_1087591026.webp)
26
i-Pulz 2
10 Errors
Any serious incident that has occurred in relation to the device should
be reported to the manufacturer.
Error-numbers
The device has an internal error detection
system. If an error is detected, the device stops
any active treatment and displays the error
screen. (See Figure 7 error screen.) Table 1 lists
the error numbers and possible solutions.
Error
Device behaviour Cause
What to do
1, 2, 3,
4, 5, 6,
9, 18
Error screen.
All functionalities
disabled.
Internal error
detected
Turn OFF the device.
When it occurs again by
turning ON, please contact
the distributor.
7, 8, 19,
20, 21
Error screen.
All functionalities
disabled.
Internal error
detected in the
output signal
Turn OFF device and
contact distributor.
Do not use the i-Pulz 2
again until it is checked
by the manufacturer or
distributor.
17
Error screen.
device turns OFF
after 10 sec
Internal error with
battery
Turn OFF device and
contact distributor.
10
Error screen
Switch to normal
mode after 10
seconds
Memory error
Press button, device ready
to use. Settings are set to
default
11
Error screen
Switch to normal
mode after 10
seconds
Memory error
Press button, device ready
to use. P13, P14, P15 are
set to default
12
Error screen
Switch to normal
mode after 10
seconds
Memory error
Press button, device ready
to use. Compliance data is
cleared
Figure 7 error screen