Tenma 72-6851 & 72-6853 Instruction manual
I? Sequential
command
Operation complete message generated immediately after <rmt> is sent.
Returns the current limit in Amps in <nr2> numeric format. The syntax of the response is
I <nr2><rmt>
Example:
If the current limit is 1·000 Amps the response to the command I? will be
I
1.000<rmt>.
OVP? Sequential
command.
Operation complete message generated immediately after <rmt> is sent.
Returns the OVP value in Volts in <nr2> numeric format. The syntax of the response is
OVP <nr2><rmt>
Example:
If the OVP value is 33.00 Volts the response to the command OVP? will be
OVP 33.00<rmt>.
VO? Sequential
command
Operation complete message generated immediately after <rmt> is sent.
Reads and returns the output voltage in Volts in <nr2> numeric format. The syntax of the
response is
<nr2>V<rmt>
Example:
If the output voltage is 12·55 Volts the response to the command VO? will be
12.55V<rmt>.
IO? Sequential
command
Operation complete message generated immediately after <rmt> is sent.
Reads and returns the output current in Amps in <nr2> numeric format. The syntax of the
response is
<nr2>A<rmt>
Example:
If the output current is 0·934 Amps the response to the command IO? will be
0.934A<rmt>
DELTAV <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Set delta voltage to <nrf>. The value of <nrf> must be in Volts; no multipliers are allowed.
If the value of <nrf>, after rounding, is outside the range of the specified output an
execution error will be generated and the corresponding error number will be placed in the
Execution Error Register, 104 (Maximum delta voltage value exceeded), if the value is too
large or 109 (Minimum delta voltage value exceeded) if the value is too small.
DELTAI <nrf>
Sequential command
Operation complete message generated immediately after execution.
Set delta Amps to <nrf>. The value of <nrf> must be in Amps; no multipliers are allowed. If
the value of <nrf>, after rounding, is outside the range of the specified output an execution
error will be generated and the corresponding error number will be placed in the
Execution Error Register, 105 (Maximum delta Amps value exceeded), if the value is too
large or 108 (Minimum delta amp value exceeded) of the value is too small.