background image

6

TEMPCO s.a. 4e ave. N°114, Z.I. des Hauts Sarts, B-4040 Herstal, Belgique. T:(32)04-2649458, F:(32)04-2649476, www.tempco.be, [email protected]  

Les photos et caractéristiques présentés dans ce catalogue sont données à titre indicatif et ne sont pas contractuelles. 

3.A.01.450-E-100910 page 3/7

Gage Mounting with Relays Remote

Where it is desirable to mount the amplifier-relay unit sepa-
rate from the gage-phototransistor unit, the gage may be
mounted either as shown in Fig. 2 (except less amplifier-
relay portion) or surface mounted as shown in Fig. 3A. Use
the layout shown in Fig. 3B to locate holes. The complete
package cannot be surface mounted.

Remote Relays Mounting

The amplifier - relay unit may be mounted remotely as
shown in Fig. 4A. Use the hole layout as shown in Fig. 4B
for this option.
Additional mounting information for special requirements is
available from the factory.

4. Pneumatic Connections & Zeroing

After installation but before making pressure connections,
set the indicating pointer exactly on the zero mark, using the
zero adjust screw located at the bottom of the front cover.
Note that this adjustment can only be made with the high
and low pressure taps both open to atmosphere.

Connect the high and low pressure taps to positive, nega-
tive, or differential pressure sensing points. Use 1/4˝ diame-
ter metal or other instrument tubing and 1/8˝ NPT adaptors
at the Dwyer Photohelic

®

pressure switch gage. Adaptors

for rubber or soft plastic tubing are furnished with the instru-
ment for use where this type of connection is preferred.

If the Photohelic

®

is not used to sense differential pressure,

one of the pressure taps must be left open to atmosphere.
This will allow the reference pressure to enter. In this case,
installation of a Dwyer No. A-331 Filter Plug or similar fitting
in the reference pressure tap is recommended to reduce the
possibility of dust entering the instrument.

NOTE:

If the Photohelic

®

switch/gage is over pressured,

pointer may ‘jump” from full scale back to zero and remain
there until the excess pressure condition is relieved. Users
should be aware of possible false zero pressure indications
under this conditions.

Fig. 3A

Remote Amplifier-Relay Unit

4-9/32

[108.74]

5

[127.00] SQ

4-9/32

[108.74]

Ø3/16

TYP 4 PLACES

Surface Mounting

Fig. 3B

Hole Layout (Front)

1-1/8

[28.58]

1-1/8

[28.58]

THREE 3/16 DIA.
HOLES AT 120

°

SURFACE MTG.

3/4 DIA. HOLE
FOR WIRE CONN.

Fig. 4A

Hole Layout

Fig. 4B

Summary of Contents for Photohelic 3001

Page 1: ...chatter free operation Series 3000 Photohelic Pressure Switch Gage Specifications Installation and Operating Instructions SPECIFICATIONS GAGE SPECIFICATIONS Service Air and non combustible compatible...

Page 2: ...be mounted remotely from the indicating gage phototransistor unit Extension cords with 7 pin plugs and receptacles are available from Dwyer for interconnection of the two units The unit may be mounte...

Page 3: ...is adjustment can only be made with the high and low pressure taps both open to atmosphere Connect the high and low pressure taps to positive nega tive or differential pressure sensing points Use 1 4...

Page 4: ...contains similar connections for the low or left set point relay The function and use of these connections varies somewhat depending on the circuit style of the instrument See paragraphs 5 and 6 for d...

Page 5: ...reset switch such as Dwyer Part No A 601 is connected to the terminals in sections B and C The circuit must be armed by momentarily closing the switch while the black pointer is to the right of the s...

Page 6: ...hus cause the low set point relay to energize If the LED should burn out only the left lower relay will de energize The right high relay will react as if pressure were above its set point and will rem...

Page 7: ...Number 3000 6MM 3000 10MM 3000 25MM 3000 50MM 3000 80MM 3000 100MM Range MM of Water 0 6 0 10 0 25 0 50 0 80 0 100 Minor Div MM 20 50 50 1 0 2 0 2 0 Model Number 3300 20MM Range Zero Center MM of Wate...

Reviews: