background image

TEMPCO s.a. 4e ave. N°114, Z.I. des Hauts Sarts, B-4040 Herstal, Belgique. T:(32)04-2649458, F:(32)04-2649476, www.tempco.be, [email protected]  

Les photos et caractéristiques présentés dans ce catalogue sont données à titre indicatif et ne sont pas contractuelles. 

3.A.01.450-E-100910 page 1/7

  

  

  

The Photohelic

®

Switch/Gage

is a very versatile, precise

pressure switch combined with the time-proven
Magnehelic

®

pressure gage. Models are available with one

or two phototransistor actuated relays. Gage reading is
unaffected by switch operation. Easy to adjust set points
have knob controls. Applied pressure and switch set points
are fully visible at all times. Deadband is one pointer width -
less than 1% of full scale. Double-pole double-throw relays
can be easily interlocked to provide variable deadband con-
trol. For positive, negative or differential pressure only on
3600S models. Full scale ranges available from 0-.25 in w.c.
to 0-6000 psig.

PHOTOHELIC

®

SENSING - HOW IT WORKS

In a typical control application, the Photohelic

®

switch/gage

controls between high and low pressure set points. When
pressure changes, reaching either set point pressure, the
beam from an LED to the limiting phototransistor will be cut
off by the helix-driven light shield. The resulting signal
change is electronically amplified to actuate its DPDT slave
relay and switching occurs. Dead band between make and
break is 1% of full scale or less - just enough to assure pos-
itive, chatter-free operation.

Series 3000 Photohelic

®

Pressure Switch/Gage 

Specifications - Installation and Operating Instructions 

SPECIFICATIONS

GAGE SPECIFICATIONS
Service: 

Air and non-combustible, compatible gases.

Wetted Materials: 

Consult Factory.

Accuracy: 

Photohelic

®

, ±2% of full scale at 70°F (21.1°C). ±3%

on -0 and ±4% on -00 models. 

Pressure Limits: 

Photohelic

®

-20" Hg. to 25 psig (-0.677 to

1.72 bar). MP  option; 35 psig (2.41 bar), HP option; 80 psig
(5.52 bar). 36003S – 36010S; 150 psig (10.34 bar). 36020S and
higher; 1.2 x full scale pressure. 

Temperature Limits: 

20 to 120°F. 

(-6.67 to 48.9°C) Low temperature option available. 

Process Connections: 

Photohelic

®

, 1/8˝ female NPT.    

Size: 

Photohelic

®

, 4˝ (101.6 mm) dial face, 5˝ (127 mm) O.D. x 8-

1/4˝ (209.55 mm). 

Weight: 

Photohelic

®

, 4 lbs. (1.81 kg). 

SWITCH SPECIFICATIONS
Switch Type: 

Each setpoint has 2 Form C relays (DPDT).

Repeatability:  

±1% of full scale.

Electrical Rating: 

Photohelic

®

,  10A @ 28 VDC, 10A @ 120,

240 VAC.

Electrical Connections: 

Screw Terminals.

Power Requirements: 

120 VAC, 50/60 Hz; 240 VAC & 24 VAC

Power optional.

Mounting Orientation: 

Diaphragm in vertical position. Consult

factory for other position orientations.

Set Point Adjustment: 

Adjustable knobs on face. 

Agency Approvals: 

Photohelic

®

UL,  CSA, CE.

2-1/2 [63.50]

2-1/16 [52.39]

2 [50.80]

1-1/4

[31.75]

(4) 6-32 HOLES

EQUALLY SPACED ON

A 5-1/8 [130.18] B.C.

Ø4-47/64

[120.25] Ø5

[127.00]

Ø4 [101.60]

FACE

5-1/2 [139.70] 

O.D.

MOUNTING 

RING

5/8 [15.88]

5/8 [15.88] PANEL 

MAX

3/16 [4.76]

3-7/8 [98.43]

5-1/8 [130.18]

6-3/8 [161.93]

(7-5/8 [193.68])

4-3/8 [111.13]

HOUSING REMOVAL

3/4 CONDUIT

CONNECTION

Ø4-3/4

[120.65]

3-7/8 SQ

[98.43]

1/8 FEMALE NPT HIGH 

PRESSURE CONNECTION

1/8 FEMALE NPT LOW

PRESSURE CONNECTION

*Patent No. 3,862,416

Summary of Contents for Photohelic 3001

Page 1: ...chatter free operation Series 3000 Photohelic Pressure Switch Gage Specifications Installation and Operating Instructions SPECIFICATIONS GAGE SPECIFICATIONS Service Air and non combustible compatible...

Page 2: ...be mounted remotely from the indicating gage phototransistor unit Extension cords with 7 pin plugs and receptacles are available from Dwyer for interconnection of the two units The unit may be mounte...

Page 3: ...is adjustment can only be made with the high and low pressure taps both open to atmosphere Connect the high and low pressure taps to positive nega tive or differential pressure sensing points Use 1 4...

Page 4: ...contains similar connections for the low or left set point relay The function and use of these connections varies somewhat depending on the circuit style of the instrument See paragraphs 5 and 6 for d...

Page 5: ...reset switch such as Dwyer Part No A 601 is connected to the terminals in sections B and C The circuit must be armed by momentarily closing the switch while the black pointer is to the right of the s...

Page 6: ...hus cause the low set point relay to energize If the LED should burn out only the left lower relay will de energize The right high relay will react as if pressure were above its set point and will rem...

Page 7: ...Number 3000 6MM 3000 10MM 3000 25MM 3000 50MM 3000 80MM 3000 100MM Range MM of Water 0 6 0 10 0 25 0 50 0 80 0 100 Minor Div MM 20 50 50 1 0 2 0 2 0 Model Number 3300 20MM Range Zero Center MM of Wate...

Reviews: