The MV2A is a thermocouple vacuum sensor controller with a range of 1
×
10
−
3
Torr to 20 Torr. It
utilizes a Televac thermocouple vacuum sensor, part number 2-2100-10 or 2-2126-0YY. It has a red,
three-digit, seven segment LED display that displays pressure in units of mTorr and Torr. The MV2A
has 2 set points relays, and a 0-5 Volt analog output.
1.3
Operating Specifications
Table 1: Operating Specifications
Operating Range
1
×
10
−
3
to 20 Torr
Communications
N/A
Programmable Set Points
2
Set Point 1
Relay
Set Point 2
Relay
Analog Output
0 to 5 V Linear
Supply Voltage
90 V to 135 V AC, 60 Hz
Calibration Medium
Dry air or nitrogen
Overpressure
Sensor-dependent
Analog Output Resolution
12 bits
Operating Temperature
0
◦
C to 50
◦
C
Storage Temperature
−
20
◦
C to 60
◦
C
Bakeout Temperature
Sensor-dependent
Response Time
≤
1 second
Accuracy
1 mTorr to 10 mTorr
±
1 mTorr
10 mTorr to 500 mTorr
±
10%
500 mTorr to 1 Torr
±
20%
Analog Output
±
10 mV
Standard Resolution
1 mTorr to 500 mTorr
1 mTorr
500 mTorr to 1 Torr
10 mTorr
1 Torr to 20 Torr
1 Torr
Display Readable Distance
8 m (26 feet)
email: [email protected]