![TelePost LP-100 Manual Download Page 23](http://html.mh-extra.com/html/telepost/lp-100/lp-100_manual_1082123023.webp)
23
Operation cont’d
Fast/Slow
This button toggles between a fast responding numerical display, a peak-hold display and starting with firmware version 1.1.47, a tune
mode. In all cases, the bar graphs remain in fast mode. The character after the numerical power readout indicates which mode you are
in. A “W” indicates peak mode, a “w” indicates fast mode and a “T” indicates tune mode. Fast mode is best for taking accurate
measurements with steady state signals, or for tuning an antenna tuner. Slow (peak) is best for CW or SSB operating. Note: The Peak
Mode is VERY fast, and can respond to a lip smack, mic button click, etc. Don’t be alarmed by this… it is normal, and allows the LP-100
to provide an accurate indication of peak power. Unless a lot of compression is used, the peak reading will be somewhat higher than
the indication with a carrier… as much as 30% depending on the ALC attack time in your rig, and power supply regulation of rig or
amplifier. Tune mode is similar to Slow mode, except that the peak hold time constant is set to 0.25 sec. The Fast and Tune modes use
the preset bargraph range in the setup section, while the Slow mode shows a fixed 13 dB range. The Tune mode locks the bargraph in
high power range.
Setup
NOTE: The screens and ordering of the screens vary slightly from version to version of the firmware. These descriptions
are based on firmware version 1.1.47b or newer.
The calibration modes can be accessed through this menu. To enter Setup mode, press Mode and Alarm in quick succession with a
little overlap. This sounds tricky, but it’s easy to master. .
NOTE: Starting with firmware version 1.1.44, entering (or exiting) Setup is
accomplished by holding the Mode button for about a second until the screen changes.
Once in Setup mode, the Mode button is used
to cycle through the calibration modes. There are 8 setup screens…
Reference.
This screen display the reference voltage from the gain/phase detector, the RSSI output from the counter AGC amplifier
and temperature in degrees F & C. It is only mainly for diagnostics. Pressing the Alarm button in this mode resets the PIC, quite useful
when flash programming the PIC. Pressing the Fast/Slow button toggles the Temp display between degrees C and F.
Dummy Load Z.
This allows you to enter the actual Z of your dummy load before calibrating. Most dummy loads are not exactly 50.0
ohms, and this will improve the accuracy of your calibration by providing the correct reference.
Gain/Phase Zero.
This screen allows band-by-band calibration of the balance of the gain and phase detectors. The process simply
requires a short transmission on each band into a dummy load, and the pressing of the Alarm/Dn button (which acts as a Save button
on this screen). The Fast/Slow/Up button becomes a Reset button in this mode, clearing the currently set corrections for the indicated
band. The LP-100 automatically gathers the correction factor and saves it indexed to frequency. The built-in frequency counter
automatically determines the frequency.
Phase Slope.
Allows calibrating the phase detector. This requires a delay line of known value. In its simplest form, this can be done by
calculating the electrical length of an existing piece of coax in the 3-10’ range, and matching the readout to the calculated length at the
frequency used for the calculation. More on this in the Calibration section. I am also working on a calibrator kit to simplify this.
Gain Slope.
Allows setting the slope of the magnitude for proper Z at a value removed from 50 ohms. This can be done with any
reasonable known load in the 25 or 75-100 ohm range. I am also working on a calibrator kit to simplify this.
Offset.
Provides for calibrating the low level ADC converter accuracy. The screen shows the output voltage of the detector, and the
Trim level is set by adjusting for zero voltage with no RF power applied.
Hi/Lo.
This screen allows the matching of the direct and divided inputs to the ADC to account for any slight variations in the precision
divider.
NOTE: This menu moved up to this position starting with version 1.1.47.
Master.
Adjusts the overall gain for power readout for all frequencies.
Fine.
Adjusts gain by band for power readout, indexed by frequency. Frequency is determined automatically by a built-in frequency
counter.
AL Thresh/Pwr Mode.
Allows selection of a power threshold for the SWR alarm. The normal setting is zero, meaning that the alarm will
work at any power level. Values of 0, 0.1, 1.0, 10.0, 100.0W. This is useful for multi-transmitter contest setups where significant energy
from a nearby antenna might be present on the output of the LP-100 coupler. If the energy is from another band, the LP-100 will display
SWR, which will be high. By setting a power threshold for the alarm, it will keep the alarm from tripping on induced power. The Pwr
Mode allows selection of Net or Fwd power. Net is Fwd-Ref… or delivered power. Fwd is the total incident power (including Ref) as
displayed on typical wattmeters like a Bird 43.
Summary of Contents for LP-100
Page 28: ...28 Schematic Page 1 ...
Page 29: ...29 Schematic Page 2 Coupler Schematic ...
Page 36: ...36 Appendix A ...