A) With a charge recovery system (preferred)
• A1) Connect the manifold’s center port to the outlet of your charge
recovery system. Connect cylinder #1 to the inlet of your charge
recovery system. Purge hoses of air.
•
A2) Open the valves on the manifold and on the cylinder. Turn on the
charge recovery system to pump the refrigerant into the Cryochiller.
•
•
A3) Repeat above steps for Tank #2 and Tank #3 of the replacement
refrigerant charge, if applicable . Stop charging when the replacement
charge cylinder is empty .
•