Performing tests 

Sprint V/V3/V4/V5

Performing tests

From the 

Main

 

Menu

 select 

Test Menu

. Refer to section 

III Operation

 for more 

details.

1. Flue gas analysis

Before performing the flue gas analysis test, check the water trap is clean and 
is upright (arrow should point in direction of gas flow). To begin test, select 

Flue 

gas  analysis

  from  the 

Test

 

Menu

Sprint  will  switch  on  the  pump 
and  begin  making  measurements. 
Check  the  fuel  type  displayed  on 
the  screen  is  correct.  Use  the  soft 
keys to log or print the results. 
Press  the  ON/OFF/ENTER  button  to 
display the three screens available: 

Screen 1: CO, CO2, CO/CO2 
ratio and pressure.
Screen 2: O2, excess air,  
temperature and efficiency.
Extra screen for V4&V5: 
O2, CO, NO, NOX
Final screen combines elements from previous screens.

To perform a pressure zero, use the soft key marked 

P=0

To end test press the ESC key.

Warning:

 During testing, ensure the combined filter and water trap is not blocked or 

full. Failure to do so may result in an error message.

ESC

14:44:27

08-Feb-2006

Flue Gas

15

Natural gas

COppm

CO2

%

Ratio

mBar

4.8

0.003

0.00

P=0

1

Pressure

Tip:

 Press the 

ON/OFF/ENTER 

button to cycle 

through the 

screens. 

Tip: 

The screen 

number is 

displayed here.

Tip: 

If # symbol 

is displayed 

- flue probe is 

not in flue or 

not connected.

Screen 1

Screen 2

Screen 3

Screen 4

Summary of Contents for Anton Sprint V2

Page 1: ...nual INS 29612 November 2011 Issue 6 Multi function flue gas analyser Tel 44 0 870 428 0075 Fax 870 428 0076 sales anton group com www anton group com 44 0 Anton Group Tel 44 0 870 428 0075 ANTON V2 2...

Page 2: ...with aggressive substances then it is the responsibility of the user to take suitable precautions that prevent it from being adversely affected 3 The equipment is designed for use in ambient tempera...

Page 3: ...overview 14 II Setting Up 16 2 1 Configuration Options 16 III Battery Charging 19 Charging the batteries 19 IV Maintenance and calibration 20 4 1 Unit 20 4 2 Water trap 20 V Specification 21 VI Access...

Page 4: ...the manual for future reference Unpacking Important ensure unit is fully charged Remove the Sprint unit from the packaging The Sprint accessories will be located in the carry case Check the contents a...

Page 5: ...o not inhale exit gas exhaust or block exit Thermocouple sockets K type connectors Flue sample inlet Display Showing start up splash screen Battery icon Three bars show complete charge ON OFF ENTER ke...

Page 6: ...e optional where suitable to the flue sam ple inlet Gas escape test Connect gas leak probe to the USB connector Note The pump may operate at different speeds or switch off depending on the test being...

Page 7: ...ure Auto zero When Sprint has completed warm up the auto zero screen will be displayed Ensure you are in clean air before proceeding Tip auto zero Sprint outside of the building or well away from the...

Page 8: ...change or ESC to cancel Units of Measurement select one of the following options Pressure units mbar Pa hPa kPa PSI inWG mmWG inHG or mmHG Temperature scale Centigrade or Fahrenheit Efficiency Net Gr...

Page 9: ...lysis a carbon monoxide alarm can be set to activate at 300ppm of CO or disabled Retrieving stored logs From the Main Menu select Stored logs and one of the following options Select a log use the soft...

Page 10: ...isplay the three screens available Screen 1 CO CO2 CO CO2 ratio and pressure Screen 2 O2 excess air temperature and efficiency Extra screen for V4 V5 O2 CO NO NOX Final screen combines elements from p...

Page 11: ...pressure source to begin pressure test 5 Press the soft key u to start test Sprint displays the duration time on the screen 6 To stop test press the soft key 7 Press the soft key to pass test and pro...

Page 12: ...est the pressure must be zeroed with the tubes con nected to the instrument but not the pressure source A thermocouple may be connected to provide temperature readings in these tests 1 Select Pressure...

Page 13: ...isplay Two probe test Connect both probes to the k type connectors See label on back of unit for FLOW and RETURN Place probes in position The screen will display the tem perature of probe 1 and probe...

Page 14: ...the test has begun Press the u soft key to start test During the test the screen will display the CO reading peak CO reading duration of test and maximum allowed CO for the test The test will run for...

Page 15: ...probe in the area of inspection for several seconds before moving it to other locations Sprint will emit continuous clicks like a Geiger counter If higher gas levels are detected the bar graph will i...

Page 16: ...urge icon will appear on the screen when purging The speed and therefore the sound emitted by the pump may vary depending on the test being conducted To save on the battery life Sprint will turn off t...

Page 17: ...n the screen These will change depending on the test or menu or if the ESC key is pressed The menu structure is shown below How to display unit serial number identity and software version Press the ES...

Page 18: ...Celsius C or degrees Fahrenheit F Pressure units are mBar Pa hPa kPa PSI inWG mmWG inHG and mmHG Efficiency units are Net Gross or NetHE Note the formulae and constants used for Gross and Net efficien...

Page 19: ...Key click settings allows you to enable or disable the key pad from making audible clicks when pressed Use the soft keys to select enabled or disabled from the submenu Press the key to accept the cha...

Page 20: ...on Due Enable or disable printing of calibration due on reports CO Alarm During flue gas analysis a carbon monoxide alarm can be set to activate at 300ppm of CO or disabled Change password Press the s...

Page 21: ...disabled The auto off timeout can be set in Analyser settings Sprint will warn user when power down is imminent Charging the batteries 1 Plug the charger into a mains socket 2 Connect the charger to...

Page 22: ...bration overdue 4 2 Water trap The combined filter and watertrap is used in line between the probe and the unit Before performing any tests check that the filter is clean and there is no water inside...

Page 23: ...ttons key pad 5 button keypad Weight 440 g 0 97lbs Pump Flow fail indication SmartPurge Enclosure Integrated robust protective case Protective rubber boot with integral magnets Standards BS7927 EN5037...

Page 24: ...en Monoxide V4 V5 only 0 1000ppm 1ppm 5ppm or 5 of reading 5ppm 90 sec 60 sec Other Measurements Measurement Range Temperature selectable C or F 50 to 1100 C 58 to 2012 F Efficiency 0 100 Net or Gross...

Page 25: ...29000 SPRINT V MAIN PROBE INC WATER TRAP WITH FILTER TUB29000 SPRINT V Neoprene Tube 1 4 ID x 3 8 OD 6 35 mm x 9 53 mm ENC29001 SPRINT V PROTECTIVE BOOT WITH MAGNETS CHG29001 SPRINT V UK CHARGER FIL29...

Page 26: ...on the screen The log data may be displayed on more than one screen Press the button to scroll through screens Use the soft key options to Print or delete the log Use the ESC key to exit the log list...

Page 27: ...printer icon will appear on the screen Press the soft key under the icon for an instant print Ensure printer is set to PC mode From software version i2 00 Printing can be aborted by a second press of...

Page 28: ...ed Check to see if charg ing symbol appears every 4 6 hours If it does not and unit does not switch on return both unit and charger for serv ice Pump sometimes oper ates at a faster rate Smart purge i...

Page 29: ...method to remotely unlock the instrument Return for reset During auto zero the gas reading does not stabilise Recovering from high gas exposure or gas sensor faulty Ensure unit is purged and allow se...

Page 30: ...it is the responsibility of the operator to ensure that the test is correctly performed to the BS 7967 2005 specification If the data does not support the result or the operator suspects it is not re...

Page 31: ...without exceeding the maximum allowed CO level and then fall by at least 1 ppm below the peak value before the end of the test Note that it is not necessary for the CO level to reach or be close to z...

Page 32: ...AIL If operator chooses to fail the test the result code is FAILED 4 otherwise the result code is PASSED 2 CO Level Exceeds Max Allowable Level for a Short Duration For some appliances eg cooker a pea...

Page 33: ...D 1 CO Level Exceeds Max Allowable Level and Peaks for Unknown Duration If the CO level exceeds the maximum allowed too close to the end of the test then the test is failed This can occur if there is...

Page 34: ...235 557700 Fax 44 0 1235 557749 Email sales telegangas co uk Web site www telegangas co uk Exclusive UK Distributor Anton Industrial Services Ltd Unit 6 Greenhill House 26 Greenhill Crescent Watford B...

Reviews: