Performing tests
Sprint V/V3/V4/V5
Performing tests
From the
Main
Menu
select
Test Menu
. Refer to section
III Operation
for more
details.
1. Flue gas analysis
Before performing the flue gas analysis test, check the water trap is clean and
is upright (arrow should point in direction of gas flow). To begin test, select
Flue
gas analysis
from the
Test
Menu
.
Sprint will switch on the pump
and begin making measurements.
Check the fuel type displayed on
the screen is correct. Use the soft
keys to log or print the results.
Press the ON/OFF/ENTER button to
display the three screens available:
Screen 1: CO, CO2, CO/CO2
ratio and pressure.
Screen 2: O2, excess air,
temperature and efficiency.
Extra screen for V4&V5:
O2, CO, NO, NOX
Final screen combines elements from previous screens.
To perform a pressure zero, use the soft key marked
P=0
To end test press the ESC key.
Warning:
During testing, ensure the combined filter and water trap is not blocked or
full. Failure to do so may result in an error message.
ESC
14:44:27
08-Feb-2006
Flue Gas
15
Natural gas
COppm
CO2
%
Ratio
mBar
4.8
0.003
0.00
P=0
1
Pressure
Tip:
Press the
ON/OFF/ENTER
button to cycle
through the
screens.
Tip:
The screen
number is
displayed here.
Tip:
If # symbol
is displayed
- flue probe is
not in flue or
not connected.
Screen 1
Screen 2
Screen 3
Screen 4