Ultra Trace Oxygen Analyzer
Operation 4
4-23
Teledyne Analytical Instruments
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. The following table
gives the range ID output for each analysis range:
Range
Voltage (V)
Current (mA)
LO
0.25
8
MED
0.50
12
HI
0.75
16
(0-1000ppm)
1.00
20