1-6
1 Introduction
Model Ultra Trace 3000
Teledyne Analytical Instruments
•
RS-232 Port
Serial digital concentration signal output
and control input.
•
Remote Probe
Used in the Ultra Trace3000 for
controlling external solenoid valves
only.
•
Remote Span/Zero
Digital inputs allow external control of
analyzer calibration.
•
Calibration Contact
To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•
Range ID Contacts
Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
•
Network I/O
Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model Ultra Trace 3000 is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.