UA-6 Detector
Section 3 Operation
3-18
3.13.2 Teledyne Isco
Retriever 500
Use cable 60-1020-217 to connect the UA-6 with a Teledyne Isco
Retriever 500 fraction collector. The information transmitted
between the two units consists of tube change and rack change
event marks.
Figure 3-12 Cable connection to the Retriever 500
3.14 Remote Chart Drive
Control by Non-Isco
Equipment
The REMOTE CHART DRIVE jack is a polarized, four-pin jack
that mates with a Cinch-Jones S-304-CCT plug or equivalent.
The pins of the plug are:
1. 50 or 60 Hz time base (at 19 VAC)
2. common
3. pulse input
4. –15 VDC
This plug can be used to stop the chart drive motor, or to drive
the motor at a variable rate.
Connect a manual switch across pins 1 and 2. The chart drive
stops when the switch is closed.
Short pins 1 and 2. The recorder chart motor may be driven with
line frequency (60 Hz at 12 to 120 VAC) across pins 3 and 4. TTL
pulses may be used by placing an optoisolator between the TTL
signal and pins 3 and 4 using the suggested circuit in Figure
3-13.
UA-6 DECTECTOR
RETRIEVER 500
CABLE #60-1020-217
Summary of Contents for UA-6
Page 2: ......
Page 4: ......
Page 12: ...x ...
Page 16: ...Isco UA 6 Table of Contents xiv ...
Page 32: ...UA 6 Detector Section 2 Preparation for Use 2 8 ...
Page 41: ...UA 6 Detector Section 3 Operation 3 9 Figure 3 6 Ferrule Application Information ...
Page 52: ...UA 6 Detector Section 3 Operation 3 20 Figure 3 14 Problem Analysis Drift and Noise ...
Page 54: ...UA 6 Detector Section 3 Operation 3 22 ...
Page 66: ...UA 6 Detector Section 4 Theory of Operation 4 12 ...
Page 90: ...UA 6 Detector Appendix A Replacement Parts A 2 ...
Page 91: ...UA 6 Detector Appendix A Replacement Parts A 3 ...
Page 92: ...UA 6 Detector Appendix A Replacement Parts A 4 MAIN CIRCUIT BOARD ...
Page 93: ...UA 6 Detector Appendix A Replacement Parts A 5 ...
Page 94: ...UA 6 Detector Appendix A Replacement Parts A 6 ...
Page 95: ...UA 6 Detector Appendix A Replacement Parts A 7 ...
Page 96: ...UA 6 Detector Appendix A Replacement Parts A 8 ...
Page 97: ...UA 6 Detector Appendix A Replacement Parts A 9 ...
Page 98: ...UA 6 Detector Appendix A Replacement Parts A 10 ...
Page 99: ...UA 6 Detector Appendix A Replacement Parts A 11 ...
Page 100: ...UA 6 Detector Appendix A Replacement Parts A 12 ...
Page 101: ...UA 6 Detector Appendix A Replacement Parts A 13 ...
Page 102: ...UA 6 Detector Appendix A Replacement Parts A 14 ...
Page 103: ...UA 6 Detector Appendix A Replacement Parts A 15 ...
Page 104: ...UA 6 Detector Appendix A Replacement Parts A 16 ...
Page 105: ...UA 6 Detector Appendix A Replacement Parts A 17 ...
Page 106: ...UA 6 Detector Appendix A Replacement Parts A 18 This page intentionally left blank ...
Page 107: ...UA 6 Detector Appendix A Replacement Parts A 19 ...
Page 108: ...UA 6 Detector Appendix A Replacement Parts A 20 ...
Page 110: ...UA 6 Detector Appendix A Replacement Parts A 22 ...
Page 111: ...UA 6 Detector Appendix A Replacement Parts A 23 ...
Page 112: ...UA 6 Detector Appendix A Replacement Parts A 24 ...
Page 114: ...UA 6 Detector Index Index 2 Lamp Replacement 5 11 Theory of Operation 4 8 Type 12 Fuses 2 4 ...
Page 116: ......
Page 118: ......
Page 120: ...Warranty ...
Page 122: ......