TIENet
®
301 pH/Temperature Device
Section 2 Installation and Setup for Signature
2-11
Figure 2-10 301 Calibration sequence
Configure Options
301 pH - pH Calibration - 4 buffer
301 pH - pH Calibration - 4 buffer
1. 4 buffer
2. 7 buffer
3. 10 buffer
4. Done
Enter exact buffer value:
pH Counts:
XXXXXXX
Select NEXT when stable
(Difference less than 10)
Place the probe in the pH buffer and
then select NEXT to continue.
Select NEXT to continue.
Perform pH Calibration of sensor
301 pH
Select pH calibration buffer or Done
pH Calibration Complete for sensor
301 pH
Place pH probe back into the flow.
pH measurements will resume in 5
minutes.
Select NEXT to continue.
Adjust Options
Level
pH (301)
pH Calibration
301 pH
X.XX
1, 2, or 3
4
1. Site Setup
2. Measurement Setup
3. Adjust
4. Equation/Trigger Setup
5. Data Storage/Push Setup
6. Sampler Setup
7. Inputs/Outputs/Alarms Setup
8. Reset Totalizers
9. Reports/History Setup
If the pH buffer is not the
ex a ct va l ue, e nte r t he
value printed on the side
of the container.
T h e s c r e e n w i l l c y c l e
repeatedly. Wait for the
n u m e r i c a l v a l u e d i s -
p l a y ed n e x t t o “ p H
counts” to stabilize to
within ±10 counts, and
then press NEXT.
Rinse the probe and submerge in
the buffer solution at least one inch
beneath the surface.
Note
The actual value displayed for “pH counts” is insignifi
-
cant to the user. It is a function of the voltage-to-fre-
quency conversion transmitted from the 301 device and
read by the Signature flow meter as a pH value.