5.  Troubleshooting 

 

Authorized Maintenance 

With proper care in installation and use, the pressure gauge will require little or no maintenance.  If 

maintenance does become necessary, it is best to consult the factory before doing so, since some procedures 

may require recalibration. Entry into the gauge or tampering with the printed circuit board will void the 

warranty. 

 

NOTICE:

 Do not perform repairs on these assemblies while the unit is still under warranty. If there is a 

problem, check the listing below to rule out some of the straight-forward solutions that may exist. 

 

 

Troubleshooting 

5.2.1.  Instrument Lists and Codes 

If a problem occurs and obvious solutions are not apparent, the internal instrument values can be accessed 

digitally. Using a terminal emulation program, type 

SFL

. This command will return a long list of values. These 

values can then be sent to the factory to better determine what the problem may be. 
Also, issues that may have occurred are recorded in memory and may be accessed using the commands 

STATUS, HISTORY

 and 

FAIL CODES

. See Commands section above. 

5.2.2.  Possible Problems and Solutions 

Symptom:

  

Not able to communicate digitally. 

Action:

   

Check settings as found in Serial Communications section above. Make certain baud rate and 

232/485 operation are correct. If using RS485 with a long cable run, make certain termination 

resistors are enabled on the last instrument on the bus. 

 

Symptom:

  

Pressure gauge is out of calibration and/or non-linear. 

Action:

   

Check all fittings for leaks. Pump system down well below 0.1 Torr and re-zero the gauge. 

 

Symptom:

 

The output is non-repeatable at zero or span (unit drifts).  

Action:

 

Make certain the instrument is properly warmed up. Pump system down well below 0.1 Torr 

and re-zero the gauge. 

 

Symptom:

 

Intermittent, “jumpy” or “noisy” output from instrument.  

Action: 

Determine if something could cause the actual pressure to fluctuate. Ensure that the 

instrument is well grounded and that all electrical connections are secure. 

 

Symptom:

  

LED indication is red. 

Action:

   

Cycle power. If symptom persists, check FAIL CODES and consult the factory. 

Summary of Contents for THCD-101

Page 1: ...INSTRUCTION MANUAL THCD 101 I S O 9 0 0 1 C E R T I F I E D www teledyne hi com...

Page 2: ...tion which if not avoided could result in death or serious injury Do not proceed beyond a WARNING notice until the indicated conditions are fully understood CAUTION indicates a hazardous situation whi...

Page 3: ...2 3 3 USB Connector USB Type C 10 2 3 4 Power Input Jack 10 3 OPERATION 11 FEATURES 11 3 1 1 Analog Input 11 3 1 2 Setpoint Control 11 3 1 3 User Rezero 11 3 1 4 Adaptive Filtering 12 3 1 5 Analog Out...

Page 4: ...commands 21 4 2 8 Other commands 21 4 2 1 Command Set Summary 22 5 TROUBLESHOOTING 24 AUTHORIZED MAINTENANCE 24 TROUBLESHOOTING 24 5 2 1 Instrument Lists and Codes 24 5 2 2 Possible Problems and Solut...

Page 5: ...ruments do not require any periodic maintenance under normal operating conditions with clean gases No damage will occur from the use of moderate overpressures 25 psig The following sections contain th...

Page 6: ...cations WARNING Do not operate instruments exceeding the specifications listed below Failure to heed this warning could result in serious personal injury and or damage to the equipment Display Range 6...

Page 7: ...out Stock Length 65 431 18 inches 65 576 4 feet 65 149 8 feet 65 148 25 feet 65 160 50 feet 65 193 100 feet 65 576 SPECIAL Custom Length cable 15 Vdc Flow Instruments 15 pin D connector U Pinout Stock...

Page 8: ...101 should be installed in a clean and careful manner Take care not to drop the product and avoid impacts to prevent damage The THCD 100 can be panel mounted or used on a bench top and is a standard...

Page 9: ...eturn 13 24 VDC 14 Set Point Out VDC All other pins Not Connected If the transducer is being powered from another source e g a mains powered supply it will only be necessary to connect to the two sign...

Page 10: ...h other settings as follows 8 data bits 1 stop bit No parity No handshaking 2 3 4 Power Input Jack The panel meter display requires a 24Vdc 5 600mA supply to operate This must be a rectified and regul...

Page 11: ...the output voltage of the setpoint would be 0 5V The setpoint does have some configuration parameters that can be used to alter the function of the setpoint control and they are detailed as follows a...

Page 12: ...f readings that are taken and buffered to calculate and show the averaged reading A filter size of 0 will naturally turn off the filtering irrespective of the band setting If the buffer size is greate...

Page 13: ...e mode switch Either mode can be cancelled by switching back to Auto mode by effectively pressing the override switch twice d Main Menu Screen The Main Menu screen allows the user to select screens th...

Page 14: ...anges A fixed IP address must be used DHCP is not an available option The THCP 101 must be on the same subnet as the computer used to communicate with it since it does not use a Gateway The chosen fix...

Page 15: ...ck to Main Menu screen Select previous character when in edit mode Select next character when in edit mode Move edit field selection up Move edit field selection down Zero Go to edit mode for currentl...

Page 16: ...change the setpoint value and to set the setpoint valve position Note The setpoint is set in engineering units as configured by the range setting and cannot be set to more than this range configuratio...

Page 17: ...ettings are On so that the filtering will be fully active filtering all data all the time and Off where there is no filtering The size is the time the data is filtered over a time greater than 5 will...

Page 18: ...cates a carriage return line feed pair For a command For a query a cmd params a cmd params a response data a response where a is the address letter cmd is the command query query identified by a after...

Page 19: ...the setpoint can be retrieved by using the query spm The returned data is SP MODE x mode string where x 0 1 or 2 and mode string AUTO OPEN or CLOSED respectively e Setpoint source sps The setpoint sou...

Page 20: ...255 0 Each octet value is again capped if a value greater than 255 is entered The THCD 101 will respond and the display the following before auto restarting the system as with the IP address UNIT SHO...

Page 21: ...ring off To retrieve the current filter size use the query fls The returned data is FILTERING SIZE x sec where x number of seconds of filtering Note if current size is 0 then returned data is FILTERIN...

Page 22: ...ost of the settings for the input channel setpoint etc The exact settings provided by this command are detailed in the table below Description Type and size Input channel units string 5 char string ra...

Page 23: ...ed See Table 3 for actual settings Ethernet IP address eip eip IP octets P1 xxx xxx xxx xxx Ethernet subnet mask esm esm IP octets P1 xxx xxx xxx xxx Input channel units string uiu uiu ccccc P1 Max le...

Page 24: ...be Also issues that may have occurred are recorded in memory and may be accessed using the commands STATUS HISTORY and FAIL CODES See Commands section above 5 2 2 Possible Problems and Solutions Symp...

Page 25: ...Material Authorization RMA form RMA Forms may be obtained from the Information section of the Hastings Instruments website http www teledyne hi com inforeq htm WARNING Contaminated parts can be detri...

Page 26: ...181 122018_THCD 101 Instruction Manual Page 26 of 26 7 Drawings...

Reviews: