2 Quick Start
Safety Terms and Symbols
The following terms may appear on the instrument:
DANGER:
Direct injury or hazard may occur.
WARNING:
Potential injury or hazard may occur.
CAUTION:
Potential damage to instrument/property may occur.
The following symbols may appear on the instrument:
CAUTION
Risk of
injury or
damage.
Refer to
manual.
WARNING
Risk of
electric
shock or
burn
Earth
Ground
Terminal
Protective
Conductor
Terminal
Frame or
Chassis
Terminal
ON/
Standby
Power
Alternating
Current
Operating Environment
T
emperature:
0 °C to 40 °C
Relative Humidity:
5 to 90% RH at ≤ 30 °C
Altitude:
≤ 3048 m at ≤ 30 °C
Use indoors only.
Pollution degree 2.
Use in an operating environment where
normally only dry, non-conductive pollution occurs. Temporary
conductivity caused by condensation should be expected.
AC Power
Input Voltage & Frequency:
100-120 V at 400 Hz or
100-240 V at 50/60 Hz
Automatic AC selection.
Power Consumption:
50 W maximum
Mains Supply Connector:
CAT II per IEC/EN 61010-1:2010,
instrument intended to be supplied from the building wiring at
utilization points (socket outlets and similar).