Teledyne T200U Manual Addendum Download Page 11

 

 

 

2.

 

SPECIFICATIONS AND AGENCY 

APPROVALS 

PARAMETER SPECIFICATION 

Ranges 

(Physical Analog Output)

 

Min: 0-5 ppb Full scale 

Max 0-2000 ppb Full scale (selectable, independent NO, NO

2

, NO

X

 ranges and auto 

ranging supported)

 

Measurement Units 

ppb, ug/m

(user selectable Single, independent or auto ranges) 

Noise at Zero

1

 

<25 ppt (RMS) 

Noise at Span

1

 

<0.5% of reading (RMS ) above 5 ppb 

Lower Detectable Limit

1

 

50 ppt (RMS) 

Zero Drift

2

 

<0.1 ppb / 24 hours  

Span Drift

2

 

<0.5% of reading or 50 ppt RMS whichever is greater /  24 hrs 

Lag Time

20 seconds 

Rise / Fall Time

1, 3

  

<50 sec to 95% 

Linearity 

1% of full scale or 

 0.1 ppb whichever is greater 

Precision 

0.5% of reading above 5 ppb 

Sample Flow Rate 

1100 cm

3

/min ±10% 

Power, Analyzer 

100V-120V, 60 Hz (62W); 220V-240V, 50 Hz (58W) 

Power, Ext Pump 

110V-120V, 60 Hz (240W), 220V-240V, 50 Hz (155W) 

Analog Output Ranges 

All outputs: 10V, 5V, 1V, 0.1V (selectable)  

Three outputs convertible to 4-20 mA isolated current loop 

All ranges with 5% under/over range 

Recorder Offset 

± 10% 

Standard I/O 

1 Ethernet: 10/100Base-T 

2 RS-232 (300 – 115,200 baud)  

2 USB device ports 

8 opto-isolated digital status outputs (7 defined, 1 spare) 

6 opto-isolated digital control inputs (4 defined, 2 spare) 

4 analog outputs 

Summary of Contents for T200U

Page 1: ...OLLUTION INSTRUMENTATION TAPI 9970 CARROLL CANYON ROAD SAN DIEGO CA 92131 1106 USA Toll free Phone 800 324 5190 Phone 1 858 657 9800 Fax 1 858 657 9816 Email api sales teledyne com Website http www teledyne api com Copyright 2010 2013 06861C DCN 6670 Teledyne Advanced Pollution Instrumentation 19 February 2013 ...

Page 2: ......

Page 3: ...addendum while the questionnaire includes Appendix C in its title there is no Appendix A or Appendix B with this addendum A Repair Questionnaire is inserted at the end of this addendum while the questionnaire includes Appendix C in its title there is no Appendix A or Appendix B with this addendum REVISION HISTORY Date Rev DCN Change Summary 2013 Feb 19 C 6670 Administrative and other minor updates...

Page 4: ...ii This page intentionally left blank ...

Page 5: ... Technician Symbol All operations marked with this symbol are to be performed by qualified maintenance personnel only Electrical Ground This symbol inside the instrument marks the central safety grounding point for the instrument CAUTION This instrument should only be used for the purpose and in the manner described in this manual If you use this instrument in a manner other than that for which it...

Page 6: ...urface chaude Ne pas toucher Toucher à certaines parties de l instrument sans protection ou sans les outils appropriés pourrait entraîner des dommages aux pièces ou à l instrument Pictogramme technicien Toutes les opérations portant ce symbole doivent être effectuées uniquement par du personnel de maintenance qualifié Mise à la terre Ce symbole à l intérieur de l instrument détermine le point cent...

Page 7: ...ations and Agency Approvals 9 2 1 EPA Equivalency Designation 10 2 2 Approvals and Certifications 11 2 2 1 SAFETY 11 2 2 2 EMC 11 2 2 3 OTHER TYPE CERTIFICATIONS 11 3 Instrument Layout 13 4 Pneumatics 15 5 Test functions 17 6 Spare Parts List 19 ...

Page 8: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 vi This page intentionally left blank ...

Page 9: ...is gold plated In addition there is a pre reactor which allows the measurements of any hydrocarbon interferents in the sample stream The sample flow is 1000 cc min The T200U is equipped with a high performance pump capable of producing a reaction cell pressure of less than 5 Inch Hg absolute Because the T200U is a close derivative of the T200 the Model T200 instruction manual continues to be valid...

Page 10: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 8 This page intentionally left blank ...

Page 11: ...er 24 hrs Lag Time1 20 seconds Rise Fall Time1 3 50 sec to 95 Linearity 1 of full scale or 0 1 ppb whichever is greater Precision 0 5 of reading above 5 ppb Sample Flow Rate 1100 cm3 min 10 Power Analyzer 100V 120V 60 Hz 62W 220V 240V 50 Hz 58W Power Ext Pump 110V 120V 60 Hz 240W 220V 240V 50 Hz 155W Analog Output Ranges All outputs 10V 5V 1V 0 1V selectable Three outputs convertible to 4 20 mA is...

Page 12: ...ensions H x W x D 7 x 17 x 23 5 178mm x 432 mm x 597 mm Weight Analyzer 40 lbs 18 kg Weight Ext Pump Pack 21 lbs 9 5 kg Environmental4 Installation Category Over voltage Category II Pollution Degree 2 1 As defined by the USEPA 2 At constant temperature and voltage 3 With adaptive filter 20 ppb change 2 1 EPA Equivalency Designation Teledyne API s T200U nitrogen oxides analyzer is designated as a r...

Page 13: ...ment for measurement control and laboratory use CE 2006 95 EC Low Voltage Directive North American cNEMKO Canada CAN CSA C22 2 No 61010 1 04 NEMKO CCL US UL No 61010 1 2nd Edition 2 2 2 EMC EN 61326 1 IEC 61326 1 Class A Emissions Industrial Immunity EN 55011 CISPR 11 Group 1 Class A Emissions FCC 47 CFR Part 15B Class A Emissions CE 2004 108 EC Electromagnetic Compatibility Directive 2 2 3 OTHER ...

Page 14: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 12 This page intentionally left blank ...

Page 15: ...13 3 INSTRUMENT LAYOUT Figure 3 1 shows the internal chassis layout of the T200U ...

Page 16: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 14 Figure 3 1 T200U Internal Chassis Layout ...

Page 17: ...ects of hydrocarbons in NOx measurements In addition the T200U is equipped with a high performance pump capable of producing a reaction cell pressure of less than 5 Inch Hg absolute SAMPLE PRESSURE SENSOR VACUUM PRESSURE SENSOR FLOW PRESSURE SENSOR PCA INSTRUMENT CHASSIS PERMAPURE DRYER PREREACTOR Orifice Dia 0 004 PMT GOLD REACTION CELL PUMP SAMPLE GAS INLET NO NOX VALVE NO2 Converter O3 Destruct...

Page 18: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 16 This page intentionally left blank ...

Page 19: ...17 5 TEST FUNCTIONS Some test functions differ between the T200U and the T200 Please refer to Appendix C this addendum for the test functions in the T200U ...

Page 20: ...Teledyne API Ultra Sensitivity Model T200U NO NO2 NOx Analyzer Addendum to T200 18 This page intentionally left blank ...

Page 21: ...ares Stocking Levels for your analyzer Note Use of replacement parts other than those supplied by TAPI may result in non compliance with European standard EN 61010 1 Note Due to the dynamic nature of spare parts lists please refer to the Website or call Customer Service for most recent updates ...

Page 22: ...LTER 47MM KB 016300900 ASSY SAMPLE FILTER 47MM ANG BKT 5UM 024600100 ASSY PREREACTOR NO NOX VALVE 024930000 CD PMT NOX 037860000 ORING TEFLON RETAINING RING 47MM KB 039700100 HEATER BAND TYPE K DUAL VOLTAGE KB 040010000 ASSY FAN REAR PANEL 040030800 PCA PRESS SENSORS 2X FLOW NOX 040400000 ASSY HEATERS THERMAL SWITCH RX CELL 040410500 ASSY VACUUM MANIFOLD 040420200 ASSY O3 GEN BRK HIGH O P 04151020...

Page 23: ...EANSER ALUMINUM 059940000 OPTION SAMPLE GAS CONDITIONER NOX 062390000 ASSY MOLY GUTS w WOOL 062420200 PCA SER INTRFACE ICOP CPU E OPTION USE WITH ICOP CPU 062870000 064540000 ASSY PUMP NOX INTERNAL 115V 60HZ 064540100 ASSY PUMP NOX INTERNAL 230V 60HZ 064540200 ASSY PUMP NOX INTERNAL 230V 50HZ 066970000 PCA INTRF LCD TOUCH SCRN F P 067240000 CPU PC 104 VSX 6154E ICOP 067300000 PCA AUX I O BD ETHERN...

Page 24: ...NG 2 006VT KB OR0000002 ORING 2 023V OR0000025 ORING 2 133V OR0000027 ORING 2 042V OR0000034 ORING 2 011V FT10 OR0000039 ORING 2 012V OR0000044 ORING 2 125V OR0000046 ORING 2 019V OR0000083 ORING 105M 1MM W X 5 MM ID VITON OR0000086 ORING 2 006 CV 75 COMPOUND KB OR0000094 ORING 2 228V 50 DURO VITON KB PU0000047 PUMP THOMAS 2688VE44 2 HEAD 115V 60HZ PU0000048 PUMP THOMAS 2688VGH144 230V 50 60HZ CE ...

Page 25: ..._______________________________________________________ PLEASE COMPLETE THE FOLLOWING TABLE NOTE DEPENDING ON OPTIONS INSTALLED NOT ALL TEST PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT IF OPTION IS INSTALLED PARAMETER RECORDED VALUE ACCEPTABLE VALUE RANGE PPB 50 PPB TO 2 000 PPB NOx STAB PPB 0 1 PPB WITH ZERO AIR SAMPLE FLOW CM3 1 000 100 OZONE FLOW CM3 80 15 PMT SIGNAL WITH ZERO A...

Page 26: ...T TEST S HAVE YOU DONE TRYING TO SOLVE THE PROBLEM ________________________________ ___________________________________________________________________________________________________ ___________________________________________________________________________________________________ ___________________________________________________________________________________________________ ________________...

Reviews: