Figure 4-9. Jumper and Cables for Multidrop Mode
Note
If you are adding an instrument to the end of a previously configured chain, remove the shunt between
Pins 21
↔
22 of JP2 on the Multidrop/LVDS PCA in the instrument that was previously the last instrument
in the chain.
4.
Close the instrument.
5.
Referring to Figure 4-10 use straight-through DB9 male
DB9 female
cables to interconnect the host RS232 port to the first analyzer’s RS232
port; then from the first analyzer’s COM2 port to the second analyzer’s
RS232 port; from the second analyzer’s COM2 port to the third analyzer’s
RS232 port, etc., connecting in this fashion up to eight analyzers, subject
to the distance limitations of the RS-232 standard.
6.
On the rear panel of each analyzer, adjust the DCE DTE switch so that the
green and the red LEDs (RX and TX) of the COM1 connector (labeled
RS232) are both lit. (Ensure you are using the correct RS-232 cables
internally wired specifically for RS-232 communication; see Table 1-1,
“Communication Cables” and the preceding subsection, “RS-232
Connection”).
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...