Pump controller rear panel connectors
1
2
3
4
5
6
7
8
9
11
10
Table 1-8 Pump Controller Rear Panel Connectors
Connector
Description
1
PUMP D
This connector is only used during multiple pump operation. The con-
trol cable from the rear panel of the fourth pump is attached to this con-
nector.
2
PUMP C
This connector is only used during multiple pump operation. The con-
trol cable from the rear panel of the
third
pump is attached to this con-
nector.
3
PUMP B
This connector is only used during multiple pump operation. The con-
trol cable from the rear panel of the
second
pump is attached to this
connector.
4
PUMP A
This plug connects the control cable from the pump rear panel. This
connection should be secured with the thumbscrews.
IMPORTANT
: The pump A connector is the only input power connec-
tor on the rear panel of the controller. During single-pump operation,
the pump must be attached to this connector to supply power to the
controller.
WARNING: Do not connect or disconnect the control cable when
the pump is connected to the mains voltage.
5
ACCESSORY
These terminals allow connection of input and output signals (such as
analog controls and external RUN/STOP).